EDITIONEN·FR·ქართ

Accréditation Sans Frontières

International Accreditation of Healthcare Facilities

ASF Standards · Ambulatory Clinic · Standard 22

Standard 22 — Gastroenterology & Endoscopy

3 criteria · 3 non-negotiable · 0 core · Version 3.0

Criteria in this standard

22.1

Endoscope Reprocessing Follows the Multisociety Guideline, With Documented Staff Competency

Non-Negotiable

Flexible endoscope reprocessing follows the current multisociety guideline in full — cleaning before high-level disinfection, model-specific manufacturer instructions, complete documentation of every step — with staff competency specifically assessed and documented before they are permitted to perform reprocessing independently.

In plain terms: Flexible endoscopes are cleaned and disinfected following the full multisociety guideline — manual cleaning first, the manufacturer's exact instructions for each model, every step recorded — by staff whose competency is documented.

Facility category Crisis Transition Small Standard
Applicability N/A Full Full Full

Why this matters

Endoscopes have been the source of hospital outbreaks of carbapenem-resistant bacteria, hepatitis, and pseudomonas. They are complex instruments with long, narrow channels that must be brushed and flushed before high-level disinfection can work; a scope that goes into the disinfector without proper manual cleaning comes out contaminated. The multisociety guideline (ASGE/SGNA/AORN/APIC) specifies each step; manufacturer instructions specify the details for each scope model. Documentation traces every scope through every reprocessing cycle to every patient. Competency means the reprocessing staff have been trained on the specific scopes and assessed as able — not that they have been doing it for years.

What good looks like

  • The written policy matches the current multisociety guideline, including model-specific instructions.
  • Staff competency is specifically assessed and documented before independent practice.
  • Every reprocessing step is documented, creating a genuine, traceable record.

Common failure modes

  • The policy is outdated or doesn't reflect model-specific manufacturer instructions.
  • Staff perform reprocessing without documented, assessed competency.
  • Reprocessing documentation is partial or assumed rather than genuinely complete.

Worked example

In practice
An endoscopy unit with six gastroscopes and four colonoscopes processed in an automated reprocessor.
BeforeManual cleaning was 'a quick brush' before the reprocessor. Channel brushing was inconsistent. No reprocessing log linked scope to patient. Staff had learned from a colleague; no competency assessment existed. Manufacturer instructions were in a drawer, unread. A cluster of post-procedure infections had been noted but not investigated.
ActionThe multisociety guideline was adopted in full. Manufacturer instructions for each scope model were laminated at the reprocessing station. A step-by-step reprocessing record is completed per scope per cycle, including manual cleaning with brush size per channel, leak test, reprocessor cycle printout, and drying, and links the scope serial number to the patient it is next used on. All reprocessing staff completed a competency assessment on each scope model, repeated annually. Quarterly surveillance cultures of scopes.
AfterThe Monitor observed a full reprocessing cycle following the guideline, reviewed two months of per-scope records linking to patients, reviewed competency records, and reviewed surveillance cultures (all negative). Verified.

If you are starting from zero — do this first

  1. Watch a scope being reprocessed. Is every channel brushed? Is a leak test done?
  2. Find the manufacturer's instructions for each scope. Post them at the station.
  3. Start a per-scope, per-cycle record linked to patient.
  4. Assess every reprocessor's competency on every scope model.
The most common mistake: Relying on the automated reprocessor — it cannot disinfect a scope that was not properly cleaned first.

Self-assessment questions

1. Does reprocessing follow a written policy matching the current multisociety guideline, including model-specific manufacturer instructions? — A specific, current, written policy, not general infection-control practice assumed to be sufficient.
Evidence: Reprocessing policy document
2. Is staff competency specifically assessed and documented before they perform reprocessing independently? — Assessed competency before independent practice, not assumed from general training.
Evidence: Staff competency assessment record
3. Is every reprocessing step documented for every instrument, creating a genuine, traceable record? — Complete, traceable documentation, not partial or assumed record-keeping.
Evidence: Reprocessing step documentation

Common reasons for a PARTIAL answer

  • Competency was assessed at hiring but hasn't been reassessed as new endoscope models were introduced. — Model-specific competency needs to reflect the actual equipment currently in use, not only equipment used at initial training.
  • Documentation is complete for the disinfection step but less consistent for the initial cleaning step. — Cleaning before disinfection is itself a critical, distinct step the guideline specifically requires be documented.
  • The policy exists but wasn't updated when the current multisociety guideline was last revised.

Implementation plan

When What
Week 1 Review current reprocessing policy against the current multisociety guideline in full.
Week 2 Establish or verify documented competency assessment for every staff member, specific to equipment currently in use.
Week 3 Confirm complete, traceable documentation exists for every reprocessing step.
Ongoing Reassess competency whenever new endoscope models are introduced.

How the Monitor verifies this

Method What Detail
DOCUMENT Reprocessing policy review Reviews the written reprocessing policy against the current multisociety guideline.
DOCUMENT Staff competency review Reviews competency assessment records for every staff member performing reprocessing.
DOCUMENT Step documentation review Reviews reprocessing records for complete, traceable documentation of every step.

Supervisor tips

  • Ask for the actual competency assessment record for a specific staff member. — A specific, documented record is the only real evidence competency was genuinely verified, not assumed.
  • Ask whether the policy has been reviewed since the guideline was last revised. — This reveals whether the facility's practice stays genuinely current, not just historically compliant.

Evidence base

[95] Day LW, Muthusamy VR, Collins J, Kushnir VM, Sultan S, Pannala R, et al. Multisociety guideline on reprocessing flexible GI endoscopes and accessories. Gastrointest Endosc. 2021;93(1):11-33 — requires a written reprocessing policy, model-specific manufacturer training, and documented staff competency prior to independent performance of high-level disinfection or sterilization.

ASF training courses on GMJ Academy →

Foundation courses A-00 to A-03 are live. Criterion-specific modules are being developed and will link here when published.

22.2

Procedural Sedation Monitoring Follows Defined Standards

Non-Negotiable

Patients receiving procedural sedation for endoscopy are monitored using continuous pulse oximetry and capnography, with vital signs recorded at defined intervals and specific, defined discharge criteria met before release — not monitored informally or released based on general appearance alone.

In plain terms: Patients sedated for endoscopy are monitored with continuous pulse oximetry and capnography, vital signs recorded at set intervals, and discharged only when they meet written criteria.

Facility category Crisis Transition Small Standard
Applicability N/A Full Full Full

Why this matters

Sedation for endoscopy causes respiratory depression that pulse oximetry detects late — oxygen saturation falls only after the patient has already stopped breathing adequately for a minute or more. Capnography detects it immediately. Deaths during endoscopy sedation are almost always from unrecognised hypoventilation. Monitoring means: continuous oximetry and capnography from sedation to recovery; blood pressure and pulse at defined intervals; a dedicated person watching the patient, not the screen; and discharge only when written criteria — awake, oriented, stable vitals, able to walk, accompanied — are met and documented.

What good looks like

  • Continuous pulse oximetry and capnography are used for every sedated patient.
  • Vital signs are recorded at defined intervals throughout procedure and recovery.
  • Specific, defined discharge criteria are met and documented before release.

Common failure modes

  • Monitoring is periodic or informal rather than continuous and objective.
  • Vital signs are recorded only at the start and end, with gaps during the procedure.
  • Discharge is based on general appearance rather than specific, documented criteria.

Worked example

In practice
An endoscopy unit performing 3,000 sedated procedures a year.
BeforeMonitoring was pulse oximetry only, watched by the endoscopy nurse who was also assisting with the procedure. No capnography. Discharge was 'when awake enough.' A patient became severely hypoxic during a prolonged colonoscopy; the oximeter alarmed after 90 seconds of apnoea. Another was discharged, fell in the car park, and fractured her wrist.
ActionCapnography was added to every sedation. A dedicated monitoring nurse (not the endoscopy assistant) watches the patient and records vital signs every five minutes during the procedure and every 15 minutes in recovery. Written discharge criteria (modified Aldrete score ≥9, able to walk unaided, oral fluids tolerated, responsible adult present) are documented and signed. Staff completed sedation monitoring training.
AfterThe Monitor observed two procedures with capnography and a dedicated monitor, reviewed 30 records with interval vital signs and completed discharge criteria, and training records. Verified.

If you are starting from zero — do this first

  1. Is capnography used for every sedation? If not, that is the gap.
  2. Assign a dedicated monitoring nurse separate from the procedure assistant.
  3. Write discharge criteria and require documentation against them.
  4. Record vital signs at defined intervals on a chart.
The most common mistake: Relying on pulse oximetry alone — by the time saturation falls, the patient has been hypoventilating for a minute.

Self-assessment questions

1. Is continuous pulse oximetry and capnography used for every patient receiving procedural sedation? — Continuous, objective monitoring, not periodic spot-checking.
Evidence: Sedation monitoring protocol
2. Are vital signs recorded at defined intervals throughout the procedure and recovery, not only at the start and end? — Consistent, scheduled recording throughout, not concentrated at the edges.
Evidence: Vital signs recording log
3. Are specific, defined discharge criteria met and documented before a patient is released, not judged from general appearance? — A specific, validated criteria set, not a subjective impression the patient seems ready.
Evidence: Discharge criteria documentation

Common reasons for a PARTIAL answer

  • Pulse oximetry is used consistently but capnography is used only for higher-sedation-level cases. — Capnography detects respiratory compromise earlier than pulse oximetry alone, at any sedation level.
  • Vital signs recording is thorough during the procedure but less consistent during recovery. — Sedation-related complications can develop during recovery, not only during the procedure itself.
  • Discharge criteria exist but aren't consistently documented as having been checked for each patient.

Implementation plan

When What
Week 1 Review current sedation monitoring practice for continuous, objective monitoring use.
Week 2 Extend capnography use to all sedation levels if not already standard.
Week 3 Establish consistent vital signs recording throughout both procedure and recovery.
Ongoing Audit discharge criteria documentation for every sedated patient.

How the Monitor verifies this

Method What Detail
OBSERVE Monitoring practice observation Observes actual sedation monitoring for continuous pulse oximetry and capnography use.
DOCUMENT Vital signs recording review Reviews recording logs for defined-interval consistency throughout procedure and recovery.
DOCUMENT Discharge criteria review Reviews documentation confirming specific discharge criteria were met before release.

Supervisor tips

  • Observe actual sedation monitoring during a procedure if timing allows. — Direct observation reveals whether monitoring is genuinely continuous, not just described as such.
  • Ask for the specific discharge criteria used and check documentation for a recent patient. — A specific, checkable record is the real evidence discharge decisions are criteria-based, not impression-based.

Evidence base

[96] Continuous pulse oximetry and capnography monitoring during procedural sedation, combined with specific, validated discharge criteria, are established practice in ambulatory endoscopy safety guidelines for detecting sedation-related respiratory compromise.

ASF training courses on GMJ Academy →

Foundation courses A-00 to A-03 are live. Criterion-specific modules are being developed and will link here when published.

22.3

Biopsy and Polyp Specimens Are Tracked to a Confirmed Pathology Result

Non-Negotiable

Every biopsy or removed polyp specimen is tracked from collection through to a confirmed, communicated pathology result, with a specific, named process ensuring no specimen result goes unreviewed or unreported to the patient — distinct from, and more specifically tracked than, general test results.

In plain terms: Every biopsy and polyp is tracked from the endoscopy room to a confirmed pathology result and to the patient being told — with a named person responsible and no specimen unaccounted for.

Facility category Crisis Transition Small Standard
Applicability N/A Full Full Full

Why this matters

A colonoscopy polyp that is cancer, whose result sits unread or whose patient is never contacted, is a curable cancer allowed to become incurable. Endoscopy units generate many specimens; most are benign; the tracking system relaxes; the one that matters is missed. The system must log every specimen at the point of collection, confirm receipt by the lab, flag results not returned by the expected date, require clinician review within a defined time, and confirm the patient has been informed — with a named owner who checks the log weekly. This is the endoscopy version of dermatology 17.1 and the same principle applies.

What good looks like

  • A specific, dedicated specimen tracking process exists, distinct from general results.
  • The system can identify and surface a specimen whose result was never received.
  • Patient communication of the result is itself tracked, not assumed.

Common failure modes

  • Specimens are tracked only within a general test-result system, with no distinct process.
  • There is no way to identify a specimen that fell through the cracks.
  • Communication to the patient is assumed but not actually tracked.

Worked example

In practice
An endoscopy unit taking 200 specimens a month.
BeforeSpecimens were labelled and sent. Results returned to the referring doctor's tray. No log. When the Coordinator matched lab records to patient notes, six results in three months had never been reviewed; one showed adenocarcinoma in a polyp; the patient was contacted four months late.
ActionA specimen log was created: patient, date, site, number of pots, sent date, lab receipt confirmation, result date, reviewed date, patient informed date. The endoscopy nurse manager owns it and reviews it weekly. Results not received in 10 days are chased. Results not reviewed in 48 hours of receipt are escalated. Malignant results trigger same-day patient contact and MDT referral. Monthly reconciliation of lab specimen count against the log.
AfterThe Monitor reviewed the log for four months: 800 specimens, all reaching 'patient informed'; median time from result to patient contact 4 days; three malignancies referred within 24 hours. Monthly reconciliation matched. Verified.

If you are starting from zero — do this first

  1. Get the lab's list of specimens from your unit last month. Match each to a reviewed result and an informed patient.
  2. Start a specimen log with every stage dated.
  3. Name an owner and review weekly.
  4. Reconcile against lab counts monthly.
The most common mistake: Sending results to the referring doctor's tray and assuming they will be actioned — trays are where results go to be forgotten.

Self-assessment questions

1. Is there a specific, named tracking process for biopsy and polyp specimens, distinct from general test result tracking? — A dedicated process specifically for these specimens, given the distinct severity risk.
Evidence: Specimen tracking log
2. Is there a way to identify a specimen whose result was never actually received or reviewed? — An active mechanism that surfaces a gap, not one that only shows completed results.
Evidence: Uncollected result identification process
3. Is the confirmed result specifically communicated to the patient, with the communication itself tracked? — Tracked communication, not an assumption the patient was informed because the result exists in the chart.
Evidence: Patient communication tracking record

Common reasons for a PARTIAL answer

  • Tracking exists for specimens collected by the primary gastroenterologist but not consistently for those collected by covering staff. — Every specimen carries the same real risk, regardless of who performed the procedure.
  • The tracking log exists but isn't reviewed on a regular schedule to catch gaps. — A log that isn't actively reviewed provides limited real protection against a missed result.
  • Results are tracked internally but patient communication specifically isn't logged.

Implementation plan

When What
Week 1 Review current specimen tracking practice for a specific, dedicated process.
Week 2 Establish a tracking log covering every specimen from collection to confirmed, communicated result.
Week 3 Build a mechanism to actively surface any specimen without a received result.
Ongoing Review the tracking log on a regular schedule for gaps.

How the Monitor verifies this

Method What Detail
DOCUMENT Specimen tracking log review Reviews the specific specimen tracking log for completeness from collection to confirmed result.
OBSERVE Uncollected result identification check Checks whether the tracking system can actually surface a specimen whose result was never received.
DOCUMENT Patient communication tracking review Reviews evidence that result communication to the patient is itself tracked, not assumed.

Supervisor tips

  • Ask for the actual tracking log and pick a specific, older specimen to trace through it. — Tracing a real, specific case reveals whether the system genuinely works, not just whether it exists.
  • Ask specifically about specimens collected during procedures performed by covering staff. — This is where tracking most commonly shows real gaps.

Evidence base

[97] Failure to track biopsy and polypectomy specimens to a confirmed, communicated pathology result is identified as a specific and serious contributor to delayed colorectal cancer diagnosis in endoscopy patient safety literature, distinct in severity from general diagnostic test follow-up.

ASF training courses on GMJ Academy →

Foundation courses A-00 to A-03 are live. Criterion-specific modules are being developed and will link here when published.

© 2026 Accréditation Sans Frontières · PHIG · Sheni Network