Standards › ASF Health Standards v10 › Standard 50
Standard 50
Diagnostic Imaging
Diagnostic imaging: radiation justified and dose optimised, equipment quality assured, reports timely with critical findings escalated, contrast media and MRI safety controlled, images retrievable, and resuscitation equipment ready..
Each criterion is shown in three settings: Humanitarian settings, Transition and fragile, and Standard. In standard settings, every criterion is Must.
Training colours: red means essential, green recommended, yellow optional. Each course shows its level for each setting: humanitarian, transition and fragile, and standard.
50.1Radiation Dose Optimisation and JustificationHumanitarian settings: MustTransition and fragile: MustStandard: Must
Every imaging study involving ionising radiation is clinically justified before ordering, with dose optimised to the lowest level consistent with diagnostic quality, not a default protocol applied regardless of patient size, age, or clinical indication.
Why it matters. Ionising radiation can harm the patient, so each scan must be justified, and the dose kept as low as the result allows. Both are required for every study.
How the surveyor verifies this criterion
Sample. Review a sample of imaging studies that use ionising radiation from the last 6 months, with their justification and dose records.
1. Every radiation study is clinically justified before it is ordered.
Method. Check each sampled study for a recorded justification dated before the study.
Evidence to show. Radiation justification record (RAD-FRM-001).
Pass when. Every sampled study has a justification recorded before it was performed.
2. Dose is optimised to the lowest level that gives diagnostic quality.
Method. Compare the dose set for each sampled study with the reference level and the protocol.
Evidence to show. Radiation justification record (RAD-FRM-001).
Pass when. Every sampled dose is within the reference level, or a reason above it is recorded.
3. Dose is audited each year, and any dose above the reference level is reviewed.
Method. Check the annual dose audit and the review of each dose above the reference level.
Evidence to show. Radiation justification and dose optimisation (RAD-POL-001).
Pass when. The annual audit is recorded and every dose above the reference level has a review.
Immediate fail (any one means the criterion is not met). A patient has a repeat CT scan within a week with no recorded justification, and the dose is never reviewed.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Nurse (ISCO 2221) | Completes the Radiation justification and dose optimisation (RAD-POL-001); radiation justification record | Must |
| Manager (ISCO 1342) | Maintains the Radiation justification and dose optimisation | Must |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| RAD-FRM-001 · Radiation justification record | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
| RAD-POL-001 · Radiation justification and dose optimisation | Chief executive (ISCO 1120) | Governing body members | N/A | Chief executive (ISCO 1120) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Diagnostic Imaging Safety and Quality Standards (AMB-16) | Manager; Nurse |
References
- IAEA (n.d.). Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards. Used for: Radiation protection and dose justification https://news.gmj.ge/guideline/radiation-protection-and-safety-of-radiation-sources-international-basic-safety-standards/
50.2Equipment Quality Assurance and CalibrationHumanitarian settings: MinimumTransition and fragile: MustStandard: Must
All imaging equipment undergoes documented quality assurance testing and calibration on a defined schedule by qualified personnel, with any equipment failing to meet standards immediately removed from clinical use until corrected.
Why it matters. An imaging machine out of calibration gives wrong pictures, and wrong pictures lead to wrong decisions. Quality testing and calibration on schedule keep the images reliable.
How the surveyor verifies this criterion
Sample. Review quality assurance records for all imaging equipment for the last 12 months.
1. All imaging equipment has documented quality assurance testing on a defined schedule.
Method. Compare the register with the schedule for each piece of equipment.
Evidence to show. Imaging equipment quality assurance register (IMG-REG-002).
Pass when. Every piece of equipment was tested within its schedule.
2. Testing and calibration are carried out by qualified personnel.
Method. Check the person who carried out each test against the qualification list.
Evidence to show. Imaging equipment quality assurance register (IMG-REG-002).
Pass when. Every test was carried out by a qualified person.
3. Equipment that fails is taken out of use until it passes.
Method. Check each failed test for the out-of-use date and the return-to-use date.
Evidence to show. Imaging equipment quality assurance and calibration (IMG-SOP-001).
Pass when. No equipment was used between a failed test and its passing retest.
Immediate fail (any one means the criterion is not met). An X-ray unit fails its calibration test, and patients are scanned on it for three weeks.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Secretary (ISCO 3344) | Keeps the Imaging equipment quality assurance (IMG-SOP-001); imaging equipment quality assurance register | Must |
| Manager (ISCO 1342) | Maintains the Imaging equipment quality assurance | Must |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| IMG-SOP-001 · Imaging equipment quality assurance and calibration | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) |
| IMG-REG-002 · Imaging equipment quality assurance register | Secretary (ISCO 3344) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Diagnostic Imaging Safety and Quality Standards (AMB-16) | Manager; Nurse |
References
- IAEA (n.d.). Radiation Protection and Safety in Medical Uses of Ionizing Radiation. Used for: Quality assurance of medical radiation equipment https://news.gmj.ge/guideline/radiation-protection-and-safety-in-medical-uses-of-ionizing-radiation/
50.3Timely, Accurate Reporting With Critical Finding EscalationHumanitarian settings: MinimumTransition and fragile: MustStandard: Must
Imaging studies are reported within a defined time frame appropriate to clinical urgency, with a documented, functioning critical finding communication protocol ensuring an urgent result reaches the ordering clinician directly, not left in a queue for routine review.
Why it matters. A critical finding on a scan, such as a bleed, must reach the clinician in time to act. A set reporting time and an escalation route for critical results protect the patient.
How the surveyor verifies this criterion
Sample. Review the reporting time and critical finding log for a sample of studies from the last 6 months, against each urgency band.
1. Studies are reported within the time frame for their clinical urgency.
Method. Compare the reporting time of each sampled study with its time frame.
Evidence to show. Imaging report turnaround and critical finding log (IMG-REG-001).
Pass when. Every sampled study was reported within its time frame, or the reason is recorded.
2. A critical finding is communicated to the responsible clinician, and the time is recorded.
Method. Check each critical finding in the log for the recipient and the time.
Evidence to show. Imaging report turnaround and critical finding log (IMG-REG-001).
Pass when. Every critical finding has a recorded recipient and time.
3. The communication protocol works.
Method. Test the protocol with a sample call to the responsible clinician and record the time to reach them.
Evidence to show. Imaging report turnaround and critical finding log (IMG-REG-001).
Pass when. The responsible clinician is reached within the time set in the protocol.
Immediate fail (any one means the criterion is not met). A critical pneumothorax is reported on a Friday evening and not communicated until Monday.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Secretary (ISCO 3344) | Keeps the Imaging report turnaround and critical finding log | Must |
| Manager (ISCO 1342) | Checks the criterion at each review | Must |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| IMG-REG-001 · Imaging report turnaround and critical finding log | Secretary (ISCO 3344) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Diagnostic Imaging Safety and Quality Standards (AMB-16) | Manager; Nurse |
References
- WHO (2021). Global Patient Safety Action Plan 2021–2030: Towards Eliminating Avoidable Harm in Health Care. Used for: Reporting and escalation of critical results https://news.gmj.ge/guideline/global-patient-safety-action-plan-2021-2030-towards-eliminating-avoidable-harm-in-health-care/
50.4Contrast Media Safety ProtocolHumanitarian settings: MinimumTransition and fragile: MediumStandard: Must
A written protocol governs contrast media use, including pre-screening for allergy and renal function, informed consent, and a documented, immediately accessible emergency response plan for a contrast reaction — not a generic allergy question treated as sufficient screening.
Why it matters. Contrast media can cause severe allergic reactions and kidney damage. A protocol that screens for allergy and kidney function before use reduces these risks.
How the surveyor verifies this criterion
Sample. Review contrast media records for a sample of contrast studies in the last 6 months, and the emergency kit checks.
1. A written protocol governs contrast media use.
Method. Check the protocol for the screening, consent and emergency steps.
Evidence to show. Contrast media safety protocol (CTR-SOP-001).
Pass when. The protocol covers allergy, renal function, consent and emergency steps.
2. Every patient is pre-screened for allergy and renal function before contrast is given.
Method. Check each sampled study for the pre-screen record before the contrast was given.
Evidence to show. Contrast media pre-screen and consent record (CTR-FRM-001).
Pass when. Every sampled study has a pre-screen recorded before the contrast was given.
3. An emergency response for reaction is documented and immediately accessible.
Method. Check the emergency kit and the daily check record, and trace the reaction response.
Evidence to show. Contrast media safety protocol (CTR-SOP-001).
Pass when. The emergency kit was checked each day and any reaction has a recorded response.
Immediate fail (any one means the criterion is not met). A patient with a known contrast allergy receives contrast, and the emergency kit is empty.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Manager (ISCO 1342) | Maintains the Contrast media safety protocol | Medium |
| Nurse (ISCO 2221) | Completes the Contrast media pre-screen and consent record | Medium |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| CTR-SOP-001 · Contrast media safety protocol | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) |
| CTR-FRM-001 · Contrast media pre-screen and consent record | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Diagnostic Imaging Safety and Quality Standards (AMB-16) | Manager; Nurse |
References
- ACR Manual on Contrast Media (American College of Radiology, 2025 (chapters updated))
50.5MRI Safety Screening and Zone ControlHumanitarian settings: MinimumTransition and fragile: MediumStandard: Must
Every patient and accompanying individual is screened for MRI contraindications before entering the magnetic field, with physical zone control preventing unscreened access — not a screening form completed but not verified before entry.
Why it matters. A metal implant or device can move or heat up in an MRI scanner and injure the patient or the person accompanying them. Screening everyone before they enter the scanner prevents that.
How the surveyor verifies this criterion
Sample. Observe MRI sessions on several days and check the screening records for a sample of studies in the last 6 months.
1. Every patient and accompanying person is screened for MRI contraindications before entry.
Method. Observe entry on several days and check the screening record for each sampled study.
Evidence to show. MRI safety screening record (MRI-FRM-001).
Pass when. Every observed and sampled entry has a screening record before entry.
2. Physical zone control prevents unscreened access to the magnetic field.
Method. Observe the zone entry on several days, and test it with an unscreened person.
Evidence to show. MRI safety screening and zone control (MRI-SOP-001).
Pass when. The zone entry stops every unscreened person, and no unscreened entry is observed.
3. Implants and devices are checked against the MRI-conditional list.
Method. Check the record of each sampled implant or device for the check.
Evidence to show. MRI safety screening record (MRI-FRM-001).
Pass when. Every sampled implant or device has a recorded check against the list.
Immediate fail (any one means the criterion is not met). A patient with an unknown metal implant enters the MRI room unscreened.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Nurse (ISCO 2221) | Completes the MRI safety screening and zone control (MRI-SOP-001); MRI safety screening record | Medium |
| Manager (ISCO 1342) | Maintains the MRI safety screening and zone control | Medium |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| MRI-SOP-001 · MRI safety screening and zone control | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) |
| MRI-FRM-001 · MRI safety screening record | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Diagnostic Imaging Safety and Quality Standards (AMB-16) | Manager; Nurse |
References
- ACR Manual on MR Safety – What's New? (American College of Radiology, 2026)
50.6Image and Report Retention, Accessible for ComparisonHumanitarian settings: MinimumTransition and fragile: MediumStandard: Must
Imaging studies and reports are retained for the required period and retrievable for comparison with prior studies, not archived in a way that makes practical retrieval difficult or unreliable.
Why it matters. Images and reports are needed for comparison, for follow-up and for the law. Keeping them for the required period, and finding them when asked, is part of safe imaging.
How the surveyor verifies this criterion
Sample. Retrieve a sample of prior imaging studies and reports from the archive, and check the retention schedule.
1. Imaging studies and reports are retained for the required period.
Method. Check the archive against the retention period for a sample of studies.
Evidence to show. Image and report retention and retrieval (ARC-POL-001).
Pass when. Every sampled study is retained for the required period.
2. Each study can be retrieved for comparison with prior studies.
Method. Retrieve a sample of studies by patient and date and time the retrieval.
Evidence to show. Image retrieval test register (ARC-REG-001).
Pass when. Every sampled study is retrieved within the set time.
3. Archiving does not make practical retrieval difficult.
Method. Check the retrieval tests for failures and any action taken.
Evidence to show. Image retrieval test register (ARC-REG-001).
Pass when. No retrieval test fails without a recorded corrective action.
Immediate fail (any one means the criterion is not met). A prior study is archived with the wrong patient identifier and cannot be retrieved for a comparison scan.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Manager (ISCO 1342) | Maintains the Image and report retention and retrieval | Medium |
| Secretary (ISCO 3344) | Keeps the Image retrieval test register | Medium |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| ARC-REG-001 · Image retrieval test register | Secretary (ISCO 3344) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
| ARC-POL-001 · Image and report retention and retrieval | Chief executive (ISCO 1120) | Governing body members | N/A | Chief executive (ISCO 1120) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Document Control and the Records System (T-02) | Manager; Nurse |
References
- Application of the Records management code of practice to radiology record retention protocols (Royal College of Radiologists, 2025)
50.7Resuscitation Equipment Is Ready Right NowHumanitarian settings: MustTransition and fragile: MustStandard: Must
Resuscitation equipment is checked every shift, fully stocked and working, and staff holding life support certification are current.
Why it matters. When a patient stops breathing, minutes matter. Equipment that is missing or broken, or staff without current life-support training, cost lives.
How the surveyor verifies this criterion
Sample. Check resuscitation equipment on every shift for a sample of days in the last 3 months, and the life support certificates of staff on those shifts.
Immediate fail (any one means the criterion is not met). The defibrillator pads are expired on the night of a cardiac arrest, and no one has checked the shift record.
Staff responsible
| Role | Responsibility | Level |
|---|---|---|
| Manager (ISCO 1342) | Checks the criterion at each review | Must |
Documents: who does each step
| Document | Writes | Approves | Fills in | Checks and signs | Files and uploads |
|---|---|---|---|---|---|
| HOS-SOP-015 · Resuscitation equipment checks | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) |
| HOS-FRM-013 · Daily resuscitation check record | Manager (ISCO 1342) | Chief executive (ISCO 1120) | Secretary (ISCO 3344) | Manager (ISCO 1342) | Secretary (ISCO 3344) |
E-learning: who must take each course
| Course | Who must take it |
|---|---|
| Emergency Medical Equipment & AED Readiness (MED-02) | Manager; Nurse |
References
- Adult basic life support Guidelines (Resuscitation Council UK (aligned with European Resuscitation Council Guidelines 2025), 2025)
Who is responsible for Standard 50
Each criterion names the positions responsible for it. Every position links to its own terms of reference.
| Criterion | Responsible | Supports |
|---|---|---|
| 50.1 Radiation Dose Optimisation and Justification | Radiation protection officer Radiologist |
Radiographer (diagnostic imaging) |
| 50.2 Equipment Quality Assurance and Calibration | Radiation protection officer Biomedical engineer |
Biomedical equipment technician |
| 50.3 Timely, Accurate Reporting With Critical Finding Escalation | Radiologist Consultant physician (internal medicine) |
Radiographer (diagnostic imaging) |
| 50.4 Contrast Media Safety Protocol | Radiologist Radiographer (diagnostic imaging) |
Head nurse / ward manager |
| 50.5 MRI Safety Screening and Zone Control | CT / MRI technologist Radiation protection officer |
Radiologist |
| 50.6 Image and Report Retention, Accessible for Comparison | Radiographer (diagnostic imaging) Clinical informatics specialist |
Quality manager |
| 50.7 Resuscitation Equipment Is Ready Right Now | Head nurse / ward manager Emergency nurse |
Quality manager |
Role details: reporting, supervision and training
Training shown is the general recommendation. Role-specific training is set in the next improvement round.
Questions and answers
The criterion text is the authority. These answers summarise it.
Who does Standard 14 apply to?
Hospitals and imaging services that perform diagnostic imaging, including X-ray, CT, MRI and contrast studies, and the resuscitation equipment kept in imaging areas.
What do the colours mean for this standard?
Red (Must) is required. Green (Medium) is recommended. Yellow (Minimum) is the lowest acceptable level in humanitarian settings. Standard setting is Must on every criterion.
Is radiation dose justified?
Yes. Under 50.1 every imaging study that uses radiation is clinically justified before it is ordered, and the dose is optimised to the lowest level that keeps diagnostic quality.
Is imaging equipment checked?
Yes. Under 50.2 all imaging equipment has documented quality assurance and calibration on a defined schedule, and faulty equipment is removed from use.
How quickly are images reported?
Under 50.3 studies are reported within a defined time for clinical urgency, and urgent results reach the responsible clinician through a working critical-finding protocol.
Are contrast media and MRI controlled?
Under 50.4 contrast media are pre-screened and consented, and under 50.5 every patient is screened for MRI contraindications before entering the magnetic field, with zone control.
Are images kept and retrievable?
Yes. Under 50.6 images and reports are kept for the required period and can be retrieved for comparison with earlier studies.
Is resuscitation equipment ready?
Under 50.7 resuscitation equipment is checked every shift and fully stocked, and staff hold current life-support certification.
Guidelines for this standard
9 guideline(s) are held in the GMJ Guidelines Library and listed here. Each link opens the full guideline on news.gmj.ge. The complete library, grouped by standard, is at Guidelines.
- 5.3 Severe dehydration — MSF Management of a cholera epidemic (Médecins Sans Frontières)
- ACR Manual on MR Safety (ACR)
- ACR Manual on MR Safety – What’s New (ACR)
- Application of the Records Management Code of Practice to Radiology Record Retention Protocols (RCR)
- ACR-AAPM-SIIM Technical Standard for Electronic Practice of Medical Imaging (American College of Radiology / AAPM / SIIM)
- ACR Manual on MR Safety (American College of Radiology)
- ACR Manual on Contrast Media (American College of Radiology)
- Diagnostic Imaging: What Is It? When and How to Use It Where Resources Are Limited? (WHO)
- Manual of diagnostic imaging: radiographic technique and projections (World Health Organization)
Also relevant to this standard:
- European Resuscitation Council Guidelines 2021: Basic Life Support (ERC)
- Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards (International Atomic Energy Agency)
- Radiation Protection and Safety in Medical Uses of Ionizing Radiation (International Atomic Energy Agency)