Standards › ASF Health Standards V11 › Standard 28

Standard 28 (Laboratory), PDF

ASF-HS11-S28 · Revision 01 · 1 November 2026 · Approved for use.

Download the PDF

Where this standard applies

Standard 28 applies to medical laboratory testing. It is required by default in hospitals, ambulatory clinics and primary health clinics, optional in long-term care, home care, dental and aesthetic clinics, and not applicable to telemedicine and fitness and wellness. The table shows each criterion and where it applies.

Criterion Hospital Ambulatory clinic Primary health clinic Long-term care Telemedicine Home care Fitness and wellness Dental clinic Aesthetic and procedure clinic
28.1 Calibration Current and Traceable to a Reference Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.2 Reagent Lots Verified Before Use Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.3 Method Validated Before Clinical Use Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.4 Internal Quality Control Run Before Results Are Released Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.5 External Quality Assessment Participation Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.6 Test Indications, Limitations and Patient Preparation Given to Users Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.7 Specimens Accepted, Logged and Rejected by Defined Criteria Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.8 Specimen Transport Preserves Integrity Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.9 STAT Specimens Prioritised and Turnaround Monitored Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.10 Chain of Custody for Forensic or Legal Specimens Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.11 Critical Values Communicated and Receipt Verified Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.12 Results Authorised Before Release Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.13 Reports Meet a Defined Minimum and Reach the Correct Recipient Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.14 Result Amendments Traceable to the Original Yes Yes Yes Yes N/A Yes N/A Yes Yes
28.15 Suppliers and Reagents Qualified Yes Yes Yes Yes N/A Yes N/A Yes Yes

Standard 28

Laboratory

It covers calibration, reagent verification, method validation, internal and external quality control, test information, specimen acceptance and transport, STAT and turnaround, chain of custody, critical values, result authorisation, reports, amendments and suppliers.

15
criteria
60
documents
5
e-learning courses
6
references

Criteria on this page

28.128.228.328.428.528.628.728.828.928.1028.1128.1228.1328.1428.15

How to read the colours
Red: Must. Required in every setting. Criteria that protect life are always red.
Yellow: Medium. Required, adapted to what a transition setting can realistically do.
Green: Minimum. The lowest acceptable level in a humanitarian setting, where resources are very limited.

Each criterion is shown in three settings: Humanitarian settings, Transition and fragile, and Standard. In standard settings, every criterion is Must.

Reference, not a requirement. Every policy, SOP, form and course on this page is a draft to help you. Adapt it to your hospital, write your own, or upload your own documents. What the assessor checks is that the standard is met and that you can show how.
Training colours: red means essential, green recommended, yellow optional. Each course shows its level for each setting: humanitarian, transition and fragile, and standard.
Calibration Current and Traceable to a Reference
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Each instrument and measuring device is calibrated or verified at the set interval against a traceable reference, and its calibration status is current before any patient result is released.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

Calibration tells the laboratory that a device reads correctly against a known reference. A device that has drifted gives results that look normal and are wrong, so the laboratory must know, for every instrument and measuring device, when it was last calibrated or verified and when it is next due.

The calibration and maintenance register (LBT-REG-001) lists each device with its interval. The calibration certificate (LBT-FRM-006) should name the reference used, so that the result can be traced back to a recognised standard. Where a device has no traceable reference, the laboratory states how it has been verified instead.

When a calibration expires, the device is taken out of use until it is calibrated again. The out-of-calibration procedure (LBT-SOP-011) says who decides this and how patient results produced in the gap are reviewed. Results from the period are not released without that review.

What good looks like

  • Each device has a calibration date, a due date and a named person responsible in the register.
  • The calibration certificate names the reference standard used.
  • Out-of-calibration devices are labelled and removed from use the same day.
  • Results produced between the last good calibration and the failure are reviewed and recorded.

Common mistakes

  • Calibration is recorded as done, but the certificate does not show traceability.
  • A device stays in use after its due date because nobody checked the register.
  • Repairs are made, but no calibration is done afterwards.
  • Verification is accepted from a supplier without checking the reference used.

Why it matters. An instrument that is out of calibration gives results that look normal and are wrong. Calibration kept current on every instrument that produces a reportable result is the basis of accurate results.

Questions the surveyor asks

  • Is every instrument and measuring device calibrated or verified at its set interval? The surveyor checks the calibration and maintenance register (LBT-REG-001) for each device in the sample against its interval.
  • Is each calibration traceable to a stated reference? The surveyor checks the calibration certificates for a traceability statement and the reference used (LBT-FRM-006).
  • Is a device taken out of use as soon as its calibration expires? The surveyor checks the out-of-calibration log and the action recorded for each event (LBT-SOP-011).

How the surveyor verifies this criterion

Sample. Review the calibration and maintenance records for all measuring devices for the last 12 months, and the traceability certificates.

1. Each device is calibrated or verified at the set interval.

Method. Check each device on the register against its calibration date and interval.

Evidence to show. Equipment calibration and traceability ( LBT-POL-001 ).

Pass when. Every device is within its calibration interval.

2. Calibration is traceable to a reference standard.

Method. Check the certificate for each calibration for the traceable reference cited.

Evidence to show. Calibration and maintenance register ( LBT-REG-001 ).

Pass when. Every calibration certificate names a traceable reference.

3. A device out of calibration is withdrawn from patient testing.

Method. Check each failed or overdue calibration for the withdrawal record.

Evidence to show. Calibration and maintenance register ( LBT-REG-001 ).

Pass when. Every failed or overdue device has a withdrawal record.

Immediate fail (any one means the criterion is not met)

  • A chemistry analyser used for patient results eight months after its calibration expired.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every device is calibrated or verified on schedule, traceable to a reference, and removed from use the moment its calibration expires.

Partial because. Calibration is recorded, but some devices are overdue or traceability is not shown on the certificates.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Equipment calibration and traceability Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-REG-001
LBT-POL-001
LBT-FRM-006
LBT-SOP-011
Secretary ISCO 3344
Keeps the Calibration and maintenance register Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-REG-001
LBT-POL-001
LBT-FRM-006
LBT-SOP-011

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-REG-001 · Calibration and maintenance register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-001 · Equipment calibration and traceability
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-006 · Calibration certificate file
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-SOP-011 · Out-of-calibration action
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
  2. International Organization for Standardization (2020). ISO 17511:2020 In vitro diagnostic medical devices: Metrological traceability of values assigned to calibrators and control materials. Geneva: ISO. https://www.iso.org/standard/69984.html
Reagent Lots Verified Before Use
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Each new reagent lot is checked against the previous lot or the manufacturer's acceptance criteria before it is used for patient testing, and the check is recorded.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A new reagent lot can behave differently from the last one, even from the same manufacturer. Each new lot is therefore checked before it is used on patients, usually by running the new lot alongside the previous lot or against controls with known values.

The verification record (LBT-FRM-001) states the acceptance criteria, the results obtained and the decision. The criteria come from the previous lot or the manufacturer's acceptance limits, and they are written down before the test is run.

A lot that fails verification is quarantined and recorded in the lot register (LBT-REG-010). It is not used on patients until the failure has been investigated and the decision to release it is signed.

What good looks like

  • Each new lot is compared with the previous lot before it is used on patients.
  • The acceptance criteria are written before the test is run.
  • A failed lot is labelled and stored separately, with the reason recorded.
  • The lot number of every reagent is recorded against each patient result.

Common mistakes

  • A new lot goes straight into use because the old one ran out.
  • The lot check is done, but the criteria are changed after the result is known.
  • Lot numbers are not recorded, so a problem cannot be traced.
  • Control results are run only on the new lot, without a comparison.

Why it matters. A new batch of reagent can behave differently from the last. Checking each new lot against the laboratory's own quality control before use prevents results going wrong unnoticed.

Questions the surveyor asks

  • Is each new reagent lot checked before it is used on patients? The surveyor checks the verification record (LBT-FRM-001) for each new lot in the sample.
  • Are the acceptance criteria for a lot defined and used? The surveyor checks the lot verification procedure (LBT-SOP-012) and the results recorded against its criteria.
  • Is a lot that fails verification kept out of use until it is resolved? The surveyor checks the lot register (LBT-REG-010) for rejected lots and the release decision recorded for each (LBT-POL-005).

How the surveyor verifies this criterion

Sample. Review the reagent lot verification records for 20 new reagent lots in the last 12 months.

1. Each new lot is checked before it is used on patients.

Method. Check each lot receipt for the verification entry dated before first use.

Evidence to show. Reagent lot verification record ( LBT-FRM-001 ).

Pass when. Every sampled lot has a verification dated before first use.

2. The check follows the manufacturer's acceptance criteria.

Method. Check each verification for the criteria and the result recorded.

Evidence to show. Reagent lot verification record ( LBT-FRM-001 ).

Pass when. Every verification records the criteria and the result.

3. A lot that fails is not used.

Method. Check each failed lot for its removal from use.

Evidence to show. Reagent lot verification record ( LBT-FRM-001 ).

Pass when. No failed lot appears in patient testing.

Immediate fail (any one means the criterion is not met)

  • A new reagent lot used on patients without any check against the previous lot.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every new reagent lot is verified against defined acceptance criteria before use, and rejected lots are quarantined and logged.

Partial because. Lots are checked, but the criteria are not defined or failed lots are not clearly kept out of use.

Staff responsible

Role Why it applies Level Training Documents
Nurse ISCO 2221
Completes the Reagent lot verification record Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-001
Manager ISCO 1342
Checks the criterion at each review Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-001
LBT-SOP-012
LBT-POL-005
LBT-REG-010

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-FRM-001 · Reagent lot verification record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-SOP-012 · Reagent lot verification procedure
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-POL-005 · Reagent and consumable acceptance
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-010 · Reagent lot register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Method Validated Before Clinical Use
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Each test method is validated for its intended clinical use, covering precision, accuracy, reportable range and reference intervals, before it is used on patients.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A test must be shown to work for the purpose it is used for. Validation covers precision, accuracy, the reportable range, the reference interval and the limits of detection and interference, as far as they apply to the test.

The validation report (LBT-FRM-007) gives the acceptance limit for each parameter and the result obtained. The report is signed before the method is used for patient care, and the method is listed in the method validation register (LBT-REG-002).

A method is validated again when it changes significantly, for example a new reagent platform, a new software version or a change in sample type. The change control policy (LBT-POL-006) says which changes need revalidation.

What good looks like

  • The validation plan is written before the study starts and states the acceptance limits.
  • Each parameter has a result, a limit and a decision in the report.
  • A method change triggers a documented review of whether revalidation is needed.
  • Method validation is repeated with a new lot of critical reagents where the manufacturer advises it.

Common mistakes

  • A method is used because the manufacturer says it is validated, without local data.
  • The report gives a result for precision but not for the reportable range.
  • The method is changed in the software but not in the validation file.
  • Validation data are kept only in the instrument, not in the register.

Why it matters. A method used for patient results must first be shown to work for this laboratory. Validation or verification before clinical use prevents results from an unproven method.

Questions the surveyor asks

  • Is each method validated for its intended clinical use before it is reported? The surveyor checks the method validation file (LBT-SOP-001) for each method in the sample.
  • Does the validation report cover precision, accuracy and the reportable range? The surveyor checks the validation report (LBT-FRM-007) for each of these parameters and the acceptance limits used.
  • Is a method revalidated after a significant change? The surveyor checks the change record for a method changed in the period and the revalidation done (LBT-POL-006, LBT-REG-002).

How the surveyor verifies this criterion

Sample. Review the method validation records for the last 5 tests introduced or changed in the last 2 years.

1. Each method is validated for its intended use before patient testing.

Method. Check each validation report for its date against the first patient result.

Evidence to show. Method validation before clinical use ( LBT-SOP-001 ).

Pass when. Every method was validated before its first patient result.

2. The validation covers precision, accuracy, reportable range and reference interval.

Method. Check each report for these four elements.

Evidence to show. Method validation register ( LBT-REG-002 ).

Pass when. Every report covers all four elements.

3. A method change triggers revalidation.

Method. Check each change in the register for a revalidation entry.

Evidence to show. Method validation register ( LBT-REG-002 ).

Pass when. Every change has a revalidation entry before use.

Immediate fail (any one means the criterion is not met)

  • A new assay put into clinical use on the manufacturer's claims alone, with no local validation.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every method is validated for its clinical use, and revalidated after each significant change, before results are reported.

Partial because. Validation is done, but some reports omit a parameter or changes are not followed by revalidation.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Method validation before clinical use Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-001
LBT-REG-002
LBT-FRM-007
LBT-POL-006
Secretary ISCO 3344
Keeps the Method validation register Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-001
LBT-REG-002
LBT-FRM-007
LBT-POL-006

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-001 · Method validation before clinical use
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-002 · Method validation register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-FRM-007 · Method validation report
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-006 · Method introduction and change control
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
  2. World Health Organization. WHO operational handbook on tuberculosis: Module 3: Diagnosis. Geneva: WHO. https://www.who.int/publications/i/item/9789240110991
Internal Quality Control Run Before Results Are Released
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Internal quality control material is run at the set interval and with each analytical run, and results are reviewed against defined rules before patient results are released.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

Internal quality control checks that the method is working on the day it is used. Control material with known values is run at the set interval and with each analytical run, and the results are compared against defined limits before patient results are released.

The rules for accepting or rejecting a run are written in the quality control policy (LBT-POL-007) and applied the same way by every staff member. A failed control stops the run, and the cause is found and corrected before the run is repeated.

Each out-of-control event is recorded on the acceptance and corrective action form (LBT-FRM-008). The record states what was done, what results were held back, and whether any patient was affected.

What good looks like

  • Control material is run at each level specified in the procedure, every day it is used.
  • The quality control rules are written down and taught to every analyst.
  • A failed control is investigated before the next patient sample is run.
  • Quality control trends are reviewed at set intervals, not only when a run fails.

Common mistakes

  • The control is run once a week but the results are reported daily.
  • A failed control is repeated until it passes, with no record of the first failure.
  • Control limits are changed without a reason recorded.
  • Results from a failed run are released because the patient needed them urgently.

Why it matters. Quality control shows whether a test is working on the day. A result released before the control has been run and accepted may be wrong.

Questions the surveyor asks

  • Is internal quality control run at the set interval and with each analytical run? The surveyor checks the internal quality control log (LBT-REG-003) against the schedule in LBT-SOP-002.
  • Are control results reviewed against defined rules before patient results are released? The surveyor checks the documented rules (LBT-POL-007) and the review record for failed controls.
  • Is each out-of-control event investigated and the corrective action recorded? The surveyor checks the QC acceptance and corrective action record (LBT-FRM-008) for each event in the sample.

How the surveyor verifies this criterion

Sample. Review the internal quality control log for 30 analytical runs and the rules applied.

1. Quality control is run at the set interval and with each run.

Method. Check each run in the sample for the quality control entry.

Evidence to show. Internal quality control ( LBT-SOP-002 ).

Pass when. Every sampled run has quality control recorded.

2. Control results are reviewed against defined rules before release.

Method. Check each run for the rule applied and the review recorded.

Evidence to show. Internal quality control log ( LBT-REG-003 ).

Pass when. Every run records the rule review before release.

3. An out-of-control run is not released and is corrected.

Method. Check each out-of-control run for the hold and the corrective action.

Evidence to show. Internal quality control log ( LBT-REG-003 ).

Pass when. Every out-of-control run is held and corrected before results are released.

Immediate fail (any one means the criterion is not met)

  • Patient results released from a run where the control was outside its limit.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Quality control runs at the set interval with every analytical run, and every failure is investigated and corrected before results are released.

Partial because. Quality control is run, but some failures are released without review or without a recorded action.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Internal quality control Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-002
LBT-REG-003
LBT-FRM-008
LBT-POL-007
Secretary ISCO 3344
Keeps the Internal quality control log Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-002
LBT-REG-003
LBT-FRM-008
LBT-POL-007

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-002 · Internal quality control
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-003 · Internal quality control log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-FRM-008 · QC acceptance and corrective action record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-007 · Quality control policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
External Quality Assessment Participation
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

The laboratory takes part in external quality assessment for each test it performs where a scheme exists, and acts on every unsatisfactory result.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

External quality assessment compares the laboratory's results with those of other laboratories using the same test, from a neutral provider. It shows whether results agree with the wider community, which internal control cannot show.

The laboratory enrols for each test where an EQA scheme is available, and records each result in the EQA register (LBT-REG-004). Where no scheme exists, the laboratory uses another method of comparison and records it.

An unsatisfactory result is investigated and the investigation recorded (LBT-FRM-009, LBT-SOP-013). Management reviews the participation policy (LBT-POL-002) at set intervals and records the decision.

What good looks like

  • The laboratory joins an EQA scheme for each test where one exists, and keeps the enrolment record.
  • Each EQA result is reviewed by a named person within the time the scheme requires.
  • An unsatisfactory result leads to a documented root cause and corrective action.
  • EQA results are reported to management with the other quality indicators.

Common mistakes

  • EQA is joined for the easy tests only, and the difficult ones are left out.
  • A result is marked satisfactory without comparing the value with the scheme's target.
  • An unsatisfactory result is not investigated because the next round was good.
  • The scheme provider is changed without a reason recorded.

Why it matters. A laboratory that does not compare its results with others may have a problem it cannot see. External quality assessment shows how its results compare with peers.

Questions the surveyor asks

  • Does the laboratory take part in external quality assessment for each test where a scheme exists? The surveyor checks the EQA results register (LBT-REG-004) against the list of tests and the schemes available.
  • Is each unsatisfactory EQA result investigated and corrected? The surveyor checks the EQA result record (LBT-FRM-009) and the investigation (LBT-SOP-013) for each unsatisfactory result.
  • Is EQA participation reviewed by management at set intervals? The surveyor checks the EQA participation policy (LBT-POL-002) and the management review minutes that refer to it.

How the surveyor verifies this criterion

Sample. Review the external quality assessment participation and results for the last 24 months for each test performed.

1. The laboratory takes part in an external scheme for each test where one exists.

Method. Check the participation record against the list of tests performed.

Evidence to show. External quality assessment participation ( LBT-POL-002 ).

Pass when. Every test with a scheme has participation recorded.

2. Each unsatisfactory result is investigated and acted on.

Method. Check each unsatisfactory result for the investigation and the corrective action.

Evidence to show. External quality assessment results register ( LBT-REG-004 ).

Pass when. Every unsatisfactory result has an investigation and action.

3. Results are reviewed at a set interval by the laboratory lead.

Method. Check the review entries for the set interval.

Evidence to show. External quality assessment results register ( LBT-REG-004 ).

Pass when. Every result is reviewed at the set interval.

Immediate fail (any one means the criterion is not met)

  • Three unsatisfactory external results for one test with no investigation.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. The laboratory takes part in EQA for every test where a scheme exists, and each unsatisfactory result leads to a documented investigation.

Partial because. EQA is joined for most tests, but some unsatisfactory results are not investigated.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the External quality assessment participation Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-REG-004
LBT-POL-002
LBT-FRM-009
LBT-SOP-013
Secretary ISCO 3344
Keeps the External quality assessment results register Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-REG-004
LBT-POL-002
LBT-FRM-009
LBT-SOP-013

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-REG-004 · External quality assessment results register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-002 · External quality assessment participation
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-009 · EQA result record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-SOP-013 · EQA unsatisfactory result investigation
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
  2. World Health Organization. Manual for organizing a national external quality assessment programme for health laboratories and other testing sites. Geneva: WHO. https://www.who.int/publications/i/item/9789241549677
Test Indications, Limitations and Patient Preparation Given to Users
Humanitarian settings: MinimumTransition and fragile: MediumStandard: Must

Clinicians and patients receive information on the indications, limitations and required preparation for each test, in the test catalogue or on the request form.

Humanitarian settings
Minimum
Transition and fragile
Medium
Standard
Must

Rating: Minimum in humanitarian settings; Medium in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

Clinicians need to know what a test can tell them, what it cannot, and how the patient should prepare. Patients need the same, in plain language. A test used for the wrong reason, or with a sample taken at the wrong time, gives a result that misleads.

The test information (LBT-SOP-010) and patient preparation sheets (LBT-FRM-002) are kept for every test. The update log (LBT-REG-011) records each change and the date it was communicated to clinicians.

The test information policy (LBT-POL-008) states who approves a change and how it is released. A change that alters the preparation or the sample requirement is communicated before the next request is accepted.

What good looks like

  • A one-page sheet for each test states the indication, the limit of the test and the preparation.
  • The sheet is dated and approved before it is issued.
  • Clinicians are told by a named route when a sheet changes.
  • Patients get the preparation in writing and in their own language.

Common mistakes

  • The test information is on a wall poster that nobody has updated for years.
  • A change to fasting or sample time is made without telling the clinicians.
  • The limits of a test are not mentioned, so results are over-read.
  • Information is copied from a manufacturer leaflet without adapting it to the local method.

Why it matters. A clinician who orders a test without knowing what it can and cannot tell them will misread the result. Clear written information on each test's purpose and limits is part of a safe result.

Questions the surveyor asks

  • Are indications, limitations and preparation for each test available to clinicians and patients? The surveyor checks the test information (LBT-SOP-010) and the preparation sheets (LBT-FRM-002) for each test in the sample.
  • Is the test information kept current, and are changes communicated? The surveyor checks the test information update log (LBT-REG-011) for changes in the period and the date each was communicated.
  • Does the test information policy set who approves each change? The surveyor checks the test information policy (LBT-POL-008) and the approval recorded for each change.

How the surveyor verifies this criterion

Sample. Review the test catalogue and request forms for 15 tests, and 10 request forms completed in the last 3 months.

1. Each test states its indication and limitations.

Method. Check each catalogue entry for the indication and limitation text.

Evidence to show. Test request information sheet ( LBT-FRM-002 ).

Pass when. Every catalogue entry states indication and limitation.

2. Patient preparation is stated for each test that needs it.

Method. Check each entry that needs preparation for the instructions given.

Evidence to show. Test request information sheet ( LBT-FRM-002 ).

Pass when. Every test that needs preparation states it.

3. Clinicians receive the information at the point of request.

Method. Check each completed request form for the information attached or referenced.

Evidence to show. Test information for users ( LBT-SOP-010 ).

Pass when. Every request form carries or refers to the test information.

Immediate fail (any one means the criterion is not met)

  • A fasting test requested with no instruction given to the patient.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Clinicians and patients have current information on the indications, limits and preparation for each test, and changes are communicated.

Partial because. Information exists for most tests, but some sheets are out of date or changes are not communicated.

Staff responsible

Role Why it applies Level Training Documents
Nurse ISCO 2221
Completes the Test request information sheet Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-002
Manager ISCO 1342
Maintains the Test information for users Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-010
LBT-FRM-002
LBT-POL-008
LBT-REG-011

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-010 · Test information for users
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-002 · Test request information sheet
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-008 · Test information policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-011 · Test information update log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. World Health Organization. WHO operational handbook on tuberculosis: Module 3: Diagnosis. Geneva: WHO. https://www.who.int/publications/i/item/9789240110991
Specimens Accepted, Logged and Rejected by Defined Criteria
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Specimens are accepted, logged and rejected against defined acceptance criteria, and every rejection is recorded with its reason and the notice given to the requester.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A specimen that is damaged, mislabelled, unsuitable or past its stability limit gives an unreliable result. The laboratory defines the acceptance criteria for each specimen type and checks each specimen on receipt against them.

Each rejection is recorded in the rejection log (LBT-REG-005) with the reason. The requesting clinician is told, using the rejection notice (LBT-FRM-010), so that a new specimen can be taken where needed.

The acceptance procedure (LBT-SOP-003) and the acceptance policy (LBT-POL-009) apply equally to every specimen. A specimen is not accepted because it is urgent, unless the criteria have been met.

What good looks like

  • The acceptance criteria are on the receiving bench, not only in a file.
  • Every rejected specimen is recorded before the sample is discarded.
  • The clinician receives a notice on the same day as the rejection.
  • Rejection rates by ward or clinic are reviewed each quarter.

Common mistakes

  • A damaged specimen is run anyway because the clinician is waiting.
  • A rejection is recorded only verbally.
  • Acceptance criteria are not defined for some specimen types.
  • The rejection reason is 'unsuitable' with nothing more written.

Why it matters. A specimen that is haemolysed, clotted or too small can give a wrong result. Written rejection criteria, applied the same way by everyone, stop it being tested.

Questions the surveyor asks

  • Are specimens accepted or rejected against defined acceptance criteria? The surveyor checks the acceptance procedure (LBT-SOP-003) and the receipt log for a sample of specimens.
  • Is each rejection recorded with the reason? The surveyor checks the rejection log (LBT-REG-005) for the reason recorded for each rejected specimen.
  • Is the requesting clinician told when a specimen is rejected? The surveyor checks the rejection notice (LBT-FRM-010) sent for each rejection in the sample.

How the surveyor verifies this criterion

Sample. Review the specimen receipt and rejection log for 30 rejected specimens and 30 accepted specimens in the last 3 months.

1. Each specimen is checked against the acceptance criteria on receipt.

Method. Check each receipt entry for the criteria checked.

Evidence to show. Specimen acceptance and rejection ( LBT-SOP-003 ).

Pass when. Every receipt records the criteria checked.

2. Each rejection records the reason and the notice given to the requester.

Method. Check each rejection for the reason and the notice date.

Evidence to show. Specimen receipt and rejection log ( LBT-REG-005 ).

Pass when. Every rejection records reason and notice.

3. Accepted specimens are logged and identified.

Method. Check each accepted specimen for the log entry and identifiers.

Evidence to show. Specimen receipt and rejection log ( LBT-REG-005 ).

Pass when. Every accepted specimen is logged with its identifiers.

Immediate fail (any one means the criterion is not met)

  • A specimen with no patient identifier accepted and tested.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Each specimen is checked against defined acceptance criteria, and every rejection is logged with its reason and notified to the clinician.

Partial because. Criteria are defined, but some rejections are not logged or the clinician is not told.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Specimen acceptance and rejection Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-003
LBT-REG-005
LBT-FRM-010
LBT-POL-009
Secretary ISCO 3344
Keeps the Specimen receipt and rejection log Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-003
LBT-REG-005
LBT-FRM-010
LBT-POL-009

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-003 · Specimen acceptance and rejection
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-005 · Specimen receipt and rejection log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-FRM-010 · Specimen rejection notice
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-009 · Specimen acceptance policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
  2. World Health Organization (2020). Laboratory biosafety manual, 4th edition. Geneva: WHO. https://www.who.int/publications/i/item/9789240011311
Specimen Transport Preserves Integrity
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Specimens are transported in conditions that preserve their integrity, including packaging, temperature, time limits and labelling, and every deviation is recorded.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

Specimens change during transport. Temperature, time, light, shaking and packaging all affect them. The laboratory defines the conditions for each specimen type and makes sure the packaging and the courier keep the specimen within them.

The transport policy (LBT-POL-010) sets the packaging standard, including the packaging used for infectious substances under the WHO Laboratory Biosafety Manual. Each shipment is recorded on the transport record (LBT-FRM-003), with the temperature and time in transit kept in the courier log (LBT-REG-012).

A specimen that has been outside its limits in transit is rejected or flagged, and the decision is recorded. The laboratory tells the sender why, so that the cause can be corrected.

What good looks like

  • The packaging meets the WHO Laboratory Biosafety Manual requirements for the specimen class.
  • Each shipment has a temperature record that is checked on arrival.
  • A courier is qualified before it is used, and the qualification is recorded.
  • A specimen outside its limits in transit is flagged on the report.

Common mistakes

  • Specimens are sent in a bag with ice that has melted.
  • Temperature loggers are fitted but the data are never read.
  • A courier is used without anyone checking how it carries specimens.
  • The transit time is not recorded, so stability cannot be judged.

Why it matters. Some specimens change in transit if they are too warm, too cold, or left too long. The transport conditions must keep the specimen as it was taken.

Questions the surveyor asks

  • Are specimens packaged and transported to preserve their integrity? The surveyor checks the transport policy (LBT-POL-010) and the packaging used for a sample of shipments.
  • Is temperature and time in transit recorded for each shipment? The surveyor checks the transport record (LBT-FRM-003) and the courier and temperature log (LBT-REG-012).
  • Are specimens that exceed their stability limits in transit rejected? The surveyor checks the rejection decisions for shipments that were outside the limits recorded.

How the surveyor verifies this criterion

Sample. Observe specimen transport from collection to the laboratory for 10 deliveries and review the transport records.

1. Specimens are packed and labelled to preserve integrity.

Method. Observe the packaging and labelling of sampled deliveries.

Evidence to show. Specimen transport ( LBT-SOP-004 ).

Pass when. Every sampled delivery is packed and labelled correctly.

2. Temperature and time limits are kept and recorded.

Method. Check each transport record for temperature and time against the limit.

Evidence to show. Specimen transport record ( LBT-FRM-003 ).

Pass when. Every record is within its temperature and time limit.

3. Each deviation is recorded and the specimen is assessed.

Method. Check each deviation for the record and the decision on the specimen.

Evidence to show. Specimen transport record ( LBT-FRM-003 ).

Pass when. Every deviation has a record and a decision.

Immediate fail (any one means the criterion is not met)

  • A specimen left in a hot vehicle for three hours and tested without a deviation record.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Specimens are packed, labelled and transported within their stability limits, with the temperature and time recorded for every shipment.

Partial because. Packaging is in place, but temperature or transit time is not recorded for some shipments.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Specimen transport Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-004
LBT-FRM-003
LBT-POL-010
LBT-REG-012
Nurse ISCO 2221
Completes the Specimen transport record Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-003

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-004 · Specimen transport
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-003 · Specimen transport record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-010 · Specimen transport policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-012 · Courier and temperature log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. World Health Organization (2020). Laboratory biosafety manual, 4th edition. Geneva: WHO. https://www.who.int/publications/i/item/9789240011311
  2. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
STAT Specimens Prioritised and Turnaround Monitored
Humanitarian settings: MinimumTransition and fragile: MediumStandard: Must

STAT specimens are identified and prioritised, and turnaround time for STAT and routine tests is monitored against defined targets.

Humanitarian settings
Minimum
Transition and fragile
Medium
Standard
Must

Rating: Minimum in humanitarian settings; Medium in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

Some results are needed within minutes to change a patient's care. The laboratory identifies STAT specimens on receipt, runs them first and tracks how long each test takes from receipt to result.

Turnaround time is recorded in the monitoring log (LBT-REG-006) for STAT and routine tests, against a target set in the priority and turnaround policy (LBT-POL-011). Targets are reviewed when they are repeatedly missed.

A STAT request is recorded on the request form (LBT-FRM-011) with the clinical reason and the requester. A request without a clinical reason is not automatically treated as STAT; the policy says how this is decided.

What good looks like

  • STAT specimens are labelled with a distinct mark and placed at the front of the queue.
  • Turnaround is measured from receipt to authorised result for each test.
  • The target for each test is written in the priority policy.
  • Missed targets are reviewed at the monthly quality meeting.

Common mistakes

  • Every request is marked STAT, so the mark means nothing.
  • The clock starts at result entry, so the delay before receipt is hidden.
  • A STAT result is released without authorisation to meet the target.
  • Targets are set once and never reviewed.

Why it matters. An urgent (STAT) specimen that waits in the queue can delay treatment for a critical patient. A written procedure to flag, handle and track urgent specimens gets them to the front.

Questions the surveyor asks

  • Are STAT specimens identified and prioritised on receipt? The surveyor checks the STAT request records (LBT-FRM-011) and the priority marking on the sampled specimens.
  • Is turnaround time for STAT and routine tests monitored against the target? The surveyor checks the turnaround time monitoring log (LBT-REG-006) for the period.
  • Is the priority policy followed and reviewed? The surveyor checks the priority and turnaround policy (LBT-POL-011) and the review record.

How the surveyor verifies this criterion

Sample. Review the STAT specimen log and the turnaround time records for the last 3 months.

1. STAT specimens are identified and prioritised on receipt.

Method. Check each STAT specimen for its identification and its priority in the workflow.

Evidence to show. STAT specimens and turnaround ( LBT-SOP-005 ).

Pass when. Every STAT specimen is identified and prioritised.

2. Turnaround is measured against the defined targets.

Method. Check the turnaround log for each target against the actual time.

Evidence to show. Turnaround time monitoring log ( LBT-REG-006 ).

Pass when. Turnaround is measured for every test against its target.

3. Targets missed are reviewed and acted on.

Method. Check each missed target for the review and action.

Evidence to show. Turnaround time monitoring log ( LBT-REG-006 ).

Pass when. Every missed target has a review and action.

Immediate fail (any one means the criterion is not met)

  • A STAT potassium result left in the queue for four hours with no alert.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. STAT specimens are identified and prioritised on receipt, and turnaround time is monitored for STAT and routine tests against a target.

Partial because. STAT specimens are marked, but turnaround is not monitored against a target for all tests.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the STAT specimens and turnaround Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-005
LBT-REG-006
LBT-POL-011
LBT-FRM-011
Secretary ISCO 3344
Keeps the Turnaround time monitoring log Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-005
LBT-REG-006
LBT-POL-011
LBT-FRM-011

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-005 · STAT specimens and turnaround
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-006 · Turnaround time monitoring log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-011 · Priority and turnaround policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-011 · STAT request record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Chain of Custody for Forensic or Legal Specimens
Humanitarian settings: MinimumTransition and fragile: MediumStandard: Must

Forensic and legal specimens are handled under a documented chain of custody from collection to disposal, and every transfer is signed.

Humanitarian settings
Minimum
Transition and fragile
Medium
Standard
Must

Rating: Minimum in humanitarian settings; Medium in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A forensic or legal specimen may be used as evidence. The laboratory must show who had it, when, and what was done to it, at every step from collection to disposal. A gap in that record can make the result unusable.

The chain of custody procedure (LBT-SOP-006) sets out the steps. The form (LBT-FRM-004) records each transfer with the name, the time and the signature of the person receiving the specimen, and the register (LBT-REG-013) lists each case.

The policy for forensic and legal specimens (LBT-POL-012) states how the specimen is stored and when it is disposed of. The disposal is recorded in the register and signed.

What good looks like

  • The custody form is started at the point of collection and travels with the specimen.
  • Every signature includes a name, a role and the time.
  • Storage locations are recorded for each specimen in the chain.
  • Disposal is recorded and witnessed.

Common mistakes

  • The custody form is filled in afterwards from memory.
  • A specimen is left unattended on a bench between transfers.
  • The name of the receiver is missing from a transfer.
  • The case is closed without recording how the specimen was disposed of.

Why it matters. A specimen used as legal evidence must show, at each step, who held it and what happened to it. Without that custody record, the evidence can be challenged.

Questions the surveyor asks

  • Is every forensic or legal specimen handled under a documented chain of custody? The surveyor checks the chain of custody procedure (LBT-SOP-006) for each forensic or legal case in the sample.
  • Is each custody transfer recorded, with the name and time of each person? The surveyor checks the chain of custody form (LBT-FRM-004) for every transfer in the sample.
  • Is the disposal of the specimen recorded? The surveyor checks the chain of custody register (LBT-REG-013) for the disposal entry of each closed case.

How the surveyor verifies this criterion

Sample. Review the chain of custody forms for 10 forensic or legal specimens in the last 12 months.

1. Each specimen has a chain of custody from collection.

Method. Check each form for the first entry at collection.

Evidence to show. Chain of custody for forensic and legal specimens ( LBT-SOP-006 ).

Pass when. Every sampled specimen has a form from collection.

2. Each transfer is signed and dated.

Method. Check each transfer on the form for a signature and date.

Evidence to show. Chain of custody form ( LBT-FRM-004 ).

Pass when. Every transfer is signed and dated.

3. The specimen is disposed of only as recorded.

Method. Check the disposal entry against the form.

Evidence to show. Chain of custody form ( LBT-FRM-004 ).

Pass when. Every disposal matches the recorded custody.

Immediate fail (any one means the criterion is not met)

  • A legal specimen with a gap in custody between two staff members.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every forensic or legal specimen has an unbroken chain of custody from collection to disposal, with each transfer recorded.

Partial because. Custody forms are used, but some transfers or the disposal of a specimen are not recorded.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Chain of custody Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-006
LBT-FRM-004
LBT-POL-012
LBT-REG-013
Nurse ISCO 2221
Completes the Chain of custody form Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-004

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-006 · Chain of custody for forensic and legal specimens
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-004 · Chain of custody form
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-012 · Forensic and legal specimen policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-013 · Chain of custody register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Critical Values Communicated and Receipt Verified
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Critical values are communicated to a responsible clinician by a defined route within a defined time, and receipt is verified and recorded.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A critical value is a result so far outside the normal range that it needs immediate action. The laboratory defines which results are critical and makes sure a responsible clinician is told within the set time.

The critical value policy (LBT-POL-013) lists the values and the time limit. The communication procedure (LBT-SOP-007) sets the route, including what happens when the responsible clinician cannot be reached. The notification record (LBT-FRM-012) and the log (LBT-REG-007) show each call.

Receipt is acknowledged by the clinician, who gives their name and the time. The route should be written down and tested, as part of the patient safety approach set out in the WHO Global Patient Safety Action Plan 2021–2030.

What good looks like

  • The list of critical values is posted at each bench and held in the information system.
  • The communication time is recorded for each notification.
  • A read-back of the value is required from the clinician.
  • The route is tested at least once a year.

Common mistakes

  • Critical values are passed on by a message left at the nurses' station.
  • The time of notification is not recorded.
  • No one checks whether the clinician received the call.
  • The critical list was set years ago and has not been reviewed since.

Why it matters. A critical result, such as a dangerously high potassium, must reach a responsible clinician at once. A result left in a report or on a screen can cost a life.

Questions the surveyor asks

  • Is a defined list of critical values in use? The surveyor checks the critical value policy (LBT-POL-013) against the list used in the laboratory information system.
  • Is each critical value communicated to a responsible clinician within the defined time? The surveyor checks the communication procedure (LBT-SOP-007) and the notification times in the sample.
  • Is receipt of each critical value acknowledged and recorded? The surveyor checks the critical value notification record (LBT-FRM-012) and the communication log (LBT-REG-007).

How the surveyor verifies this criterion

Sample. Review the critical value log for 30 critical results in the last 3 months.

1. Each critical value is communicated to a responsible clinician.

Method. Check each log entry for the recipient named.

Evidence to show. Critical value communication ( LBT-SOP-007 ).

Pass when. Every critical value names a responsible recipient.

2. Communication happens within the defined time.

Method. Check each entry for the time from result to communication.

Evidence to show. Critical value communication log ( LBT-REG-007 ).

Pass when. Every communication is within the defined time.

3. Receipt is verified and recorded.

Method. Check each entry for the read-back and receipt confirmation.

Evidence to show. Critical value communication log ( LBT-REG-007 ).

Pass when. Every entry records receipt confirmed.

Immediate fail (any one means the criterion is not met)

  • A critical potassium result left on a voicemail and not acted on.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. A defined list of critical values is communicated to a responsible clinician within the set time, and each notification is acknowledged and recorded.

Partial because. Critical values are phoned, but the time is not recorded or receipt is not acknowledged.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Critical value communication Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-007
LBT-REG-007
LBT-FRM-012
LBT-POL-013
Secretary ISCO 3344
Keeps the Critical value communication log Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-007
LBT-REG-007
LBT-FRM-012
LBT-POL-013

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-007 · Critical value communication
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-007 · Critical value communication log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-FRM-012 · Critical value notification record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-013 · Critical value policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. World Health Organization (2021). Global Patient Safety Action Plan 2021–2030: Towards eliminating avoidable harm in health care. Geneva: WHO. https://www.who.int/publications-detail-redirect/9789240032705
  2. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Results Authorised Before Release
Humanitarian settings: MinimumTransition and fragile: MustStandard: Must

Each result is authorised for release by a qualified person before it leaves the laboratory, and the person authorising is recorded.

Humanitarian settings
Minimum
Transition and fragile
Must
Standard
Must

Rating: Minimum in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A result is only released after a qualified person has checked it against the quality control, the patient's history and the expected range. Authorisation is the last check before the result leaves the laboratory.

The authorisation procedure (LBT-SOP-014) and the policy (LBT-POL-003) state who may authorise each type of result. The authorisation record (LBT-FRM-005) and the authorised results register (LBT-REG-014) show who released each result and when.

A result that does not pass the check is held. It is not released until the cause has been found and the result has been checked again.

What good looks like

  • The authoriser is named on the report, with their qualification and role.
  • A result that falls outside the expected range is checked against the history before it is released.
  • The authorisation is recorded before the report leaves the laboratory.
  • A result that cannot be authorised is held and the reason recorded.

Common mistakes

  • Results are released automatically without any human check, and nobody reviews the rules.
  • The authoriser is also the person who ran the test, with no second check.
  • A result is released by a person who is not qualified for that test.
  • The authorisation is given verbally and not recorded.

Why it matters. A result released without being checked by qualified staff can carry an error to the clinician. A review and authorisation before release is the last check.

Questions the surveyor asks

  • Is each result authorised by a qualified person before release? The surveyor checks the authorisation record (LBT-FRM-005) for each result in the sample.
  • Is the authorisation procedure defined and followed? The surveyor checks the result authorisation procedure (LBT-SOP-014) and the policy (LBT-POL-003).
  • Does the authorised results register show who released each result, and when? The surveyor checks the authorised results register (LBT-REG-014) for each result in the sample.

How the surveyor verifies this criterion

Sample. Review the authorisation records for 30 released results and the competence records of the authorising staff.

1. Each result is authorised before release by a qualified person.

Method. Check each result for the authoriser and the time before release.

Evidence to show. Result authorisation and release ( LBT-POL-003 ).

Pass when. Every released result has an authoriser before release.

2. The authoriser is competent for the test.

Method. Check the authoriser against the competence register for the test.

Evidence to show. Result authorisation record ( LBT-FRM-005 ).

Pass when. Every authoriser is competent for the test.

3. The authoriser is recorded with the result.

Method. Check each result record for the name recorded.

Evidence to show. Result authorisation record ( LBT-FRM-005 ).

Pass when. Every released result shows the authoriser.

Immediate fail (any one means the criterion is not met)

  • A result released by a trainee with no authorisation.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every result is authorised by a qualified person before release, and the register shows who released it and when.

Partial because. Results are authorised, but the authoriser is not always qualified or the release is not recorded.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Result authorisation and release Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-005
LBT-POL-003
LBT-SOP-014
LBT-REG-014
Nurse ISCO 2221
Completes the Result authorisation record Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-FRM-005

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-FRM-005 · Result authorisation record
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-003 · Result authorisation and release
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-SOP-014 · Result authorisation procedure
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-014 · Authorised results register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Reports Meet a Defined Minimum and Reach the Correct Recipient
Humanitarian settings: MustTransition and fragile: MustStandard: Must

Each report contains a defined minimum set of information, including identifiers, test name, result, units, reference interval and authoriser, and reaches the correct recipient.

Humanitarian settings
Must
Transition and fragile
Must
Standard
Must

Rating: Must in humanitarian settings; Must in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A report is the only part of the laboratory's work that most clinicians see. It must carry enough information to be read correctly: the patient identifier, the test name, the result, the units, the reference interval and the date and time of collection and reporting.

The report content procedure (LBT-SOP-008) defines the minimum set. The template checklist (LBT-FRM-013) is used whenever a template is changed. The release log (LBT-REG-015) records each release, and the report policy (LBT-POL-014) says how reports are delivered.

A report that is delivered in a different form, such as a fax or a message, carries the same information and is recorded in the same way.

What good looks like

  • Every report carries the same minimum information, in the same order.
  • The reference interval is printed next to each result, with its units.
  • The report shows the date and time of collection and of release.
  • Corrections are shown on the report, not hidden.

Common mistakes

  • A report has no units, so a number cannot be read correctly.
  • The reference interval is from an older method and is not updated.
  • The report omits the requesting clinician's name.
  • Reports are released in batches without a record.

Why it matters. A report without the patient's identity, the specimen type or the test date cannot be used safely. Every report must carry a defined minimum of information.

Questions the surveyor asks

  • Does each report contain the defined minimum set of information? The surveyor checks the report content procedure (LBT-SOP-008) and each sampled report against it.
  • Is the report template checked before it is used? The surveyor checks the template checklist (LBT-FRM-013) for each template in use.
  • Is each report release logged? The surveyor checks the report release log (LBT-REG-015) for each report in the sample against the report policy (LBT-POL-014).

How the surveyor verifies this criterion

Sample. Review 20 laboratory reports and the delivery records for the last 3 months.

1. Each report contains the defined minimum information.

Method. Check each report against the minimum content list.

Evidence to show. Laboratory report content and delivery ( LBT-SOP-008 ).

Pass when. Every report contains all the minimum content.

2. The report reaches the correct recipient.

Method. Check each delivery against the requester recorded.

Evidence to show. Laboratory report content and delivery ( LBT-SOP-008 ).

Pass when. Every report reaches the requester named.

3. Reports carry the authoriser and reference interval.

Method. Check each report for the authoriser and the reference interval.

Evidence to show. Laboratory report content and delivery ( LBT-SOP-008 ).

Pass when. Every report carries both.

Immediate fail (any one means the criterion is not met)

  • A report sent to the wrong ward, with another patient's result on it.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every report carries the defined minimum information, uses a checked template, and each release is logged.

Partial because. Reports are issued, but some omit units, reference ranges or the release time.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Laboratory report content and delivery Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-008
LBT-POL-014
LBT-FRM-013
LBT-REG-015
Manager ISCO 1342
Checks the criterion at each review Must Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-008
LBT-POL-014
LBT-FRM-013
LBT-REG-015

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-008 · Laboratory report content and delivery
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-POL-014 · Laboratory report policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-013 · Report template checklist
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-REG-015 · Report release log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Result Amendments Traceable to the Original
Humanitarian settings: MinimumTransition and fragile: MediumStandard: Must

Any change to a released result is recorded as an amendment, traceable to the original result, with the reason, the date and the person who made the change.

Humanitarian settings
Minimum
Transition and fragile
Medium
Standard
Must

Rating: Minimum in humanitarian settings; Medium in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

A result that is later found to be wrong must be corrected, but the original must remain visible. Any change to a released result is recorded as an amendment, so that a clinician can see what was reported first, what was changed and why.

The amendment procedure (LBT-SOP-009) and policy (LBT-POL-015) say who may authorise an amendment. The amendment form (LBT-FRM-014) states the reason and links the amended result to the original. The amendment log (LBT-REG-008) lists every change.

Where the amended result changes the clinical meaning, the clinician is told directly and the notification is recorded.

What good looks like

  • An amendment shows the original result and the corrected result, with the reason.
  • The amendment is signed by an authorised person other than the one who first released the result.
  • The amendment is logged in the amendment register.
  • The clinician is told when the change alters the clinical meaning.

Common mistakes

  • The original result is overwritten, so the change cannot be seen.
  • An amendment is made without a reason written down.
  • The amendment is made by the person who first released the result, with no check.
  • Amendments are made in the information system without a log.

Why it matters. A result that is changed after release must show what it was and what it became, and why. Without that, the change cannot be trusted or reviewed.

Questions the surveyor asks

  • Is each change to a released result recorded as an amendment? The surveyor checks the amendment log (LBT-REG-008) for each amendment in the period.
  • Does the amendment link to the original result and give the reason for the change? The surveyor checks the amendment form (LBT-FRM-014) and the link to the original result for each sampled amendment.
  • Is the amendment policy followed, including who may authorise an amendment? The surveyor checks the amendment policy (LBT-POL-015) and the authorisation recorded for each amendment in the sample (LBT-SOP-009).

How the surveyor verifies this criterion

Sample. Review the result amendment log for all amendments in the last 12 months.

1. Each amendment is recorded as an amendment to the original result.

Method. Check each amendment for the link to the original result.

Evidence to show. Result amendments ( LBT-SOP-009 ).

Pass when. Every amendment is linked to the original.

2. Each amendment states its reason, date and person.

Method. Check each amendment entry for the reason, date and person.

Evidence to show. Result amendment log ( LBT-REG-008 ).

Pass when. Every amendment states reason, date and person.

3. The original result remains visible in the record.

Method. Check the record for the original value kept.

Evidence to show. Result amendment log ( LBT-REG-008 ).

Pass when. The original value is kept visible in every sampled case.

Immediate fail (any one means the criterion is not met)

  • A released result overwritten with no record of the original value.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every change to a released result is an amendment linked to the original and signed off by an authorised person, with the reason recorded.

Partial because. Amendments are recorded, but the reason or the link to the original result is missing for some.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Result amendments Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-009
LBT-REG-008
LBT-FRM-014
LBT-POL-015
Secretary ISCO 3344
Keeps the Result amendment log Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-SOP-009
LBT-REG-008
LBT-FRM-014
LBT-POL-015

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-SOP-009 · Result amendments
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-REG-008 · Result amendment log
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-FRM-014 · Result amendment form
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-015 · Result amendment policy
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
Suppliers and Reagents Qualified
Humanitarian settings: MinimumTransition and fragile: MediumStandard: Must

Suppliers of reagents, consumables and equipment services are qualified before use and reviewed at set intervals against defined requirements.

Humanitarian settings
Minimum
Transition and fragile
Medium
Standard
Must

Rating: Minimum in humanitarian settings; Medium in transition and fragile settings; Must in standard settings.

Not assessed (N/A) for: telemedicine, fitness and wellness.

Guidance

Reagents, consumables and equipment services affect every result. A supplier that delivers a faulty product or an unqualified service can compromise many tests before anyone notices. The laboratory qualifies each supplier before first use and reviews it at a set interval.

The supplier qualification register (LBT-REG-009) lists each supplier, the products or services supplied, the qualification decision and the review date. The qualification form (LBT-FRM-015) records the evidence used, such as certificates or performance data.

The qualification procedure (LBT-SOP-015) and policy (LBT-POL-004) apply to all new suppliers, including those offered at a lower price. A supplier that fails review is removed from the approved list.

What good looks like

  • Each supplier is qualified on a defined set of evidence before its first order.
  • Supplier performance is reviewed each year, with EQA results and complaints considered.
  • A supplier that fails review is removed from the approved list.
  • Equipment service engineers are qualified and their work is checked.

Common mistakes

  • A new supplier is used because it is cheaper, without qualification.
  • A supplier is never reviewed after the first order.
  • A service engineer works on a device without a record of the work done.
  • A product is replaced by a different brand without a new qualification.

Why it matters. Reagents, consumables and outside services that are poor quality cause results to be wrong. Buying only from qualified suppliers, and monitoring them, keeps quality up.

Questions the surveyor asks

  • Are reagent, consumable and service suppliers qualified before first use? The surveyor checks the supplier qualification register (LBT-REG-009) for each supplier in the sample.
  • Is each supplier reviewed at a set interval? The surveyor checks the supplier qualification form (LBT-FRM-015) and the review date for each supplier in the sample.
  • Is the qualification procedure followed for new suppliers? The surveyor checks the qualification procedure (LBT-SOP-015) and the policy (LBT-POL-004) against the suppliers added in the period.

How the surveyor verifies this criterion

Sample. Review the supplier qualification register and the purchase records for all reagent and service suppliers in use.

1. Each supplier is qualified before first use.

Method. Check each supplier on the register for the qualification date before the first purchase.

Evidence to show. Supplier and reagent qualification ( LBT-POL-004 ).

Pass when. Every supplier was qualified before first use.

2. Qualification is reviewed at the set interval.

Method. Check each supplier for the review date against the interval.

Evidence to show. Supplier qualification register ( LBT-REG-009 ).

Pass when. Every supplier is reviewed at the set interval.

3. Unqualified suppliers are not used.

Method. Check the purchase records against the register.

Evidence to show. Supplier qualification register ( LBT-REG-009 ).

Pass when. No purchase is made from an unqualified supplier.

Immediate fail (any one means the criterion is not met)

  • Reagents bought from an unlisted supplier with no quality documents on file.

Staff, training and documents

Each occupation carries its ISCO-08 code. Each row is one document: read across to see who does each step. Training evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or your own training certificate kept in the staff file.

Good practice. Every reagent, consumable and service supplier is qualified before first use and reviewed at a set interval.

Partial because. Most suppliers are qualified, but some are used before qualification or not reviewed on time.

Staff responsible

Role Why it applies Level Training Documents
Manager ISCO 1342
Maintains the Supplier and reagent qualification Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-REG-009
LBT-POL-004
LBT-FRM-015
LBT-SOP-015
Secretary ISCO 3344
Keeps the Supplier qualification register Medium Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Evidence: a verified micro-credential in your GMJ Academy micro-credential portfolio, or a training certificate in the staff file
LBT-REG-009
LBT-POL-004
LBT-FRM-015
LBT-SOP-015

E-learning: who must take each course.

E-learning course Who must take Who is recommended
Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
GMJ Academy: English · Georgian
Manager Nurse, Secretary
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
GMJ Academy: English · Georgian
Manager, Nurse None

Documents: who does each step

Document Writes Approves Fills in Checks and signs Files and uploads
LBT-REG-009 · Supplier qualification register
Register
Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-POL-004 · Supplier and reagent qualification
Policy
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)
LBT-FRM-015 · Supplier qualification form
Form (record)
Manager (ISCO 1342) Chief executive (ISCO 1120) Nurse (ISCO 2221) Manager (ISCO 1342) Secretary (ISCO 3344)
LBT-SOP-015 · Supplier qualification procedure
Procedure (SOP)
Manager (ISCO 1342) Chief executive (ISCO 1120) Manager (ISCO 1342) Chief executive (ISCO 1120) Secretary (ISCO 3344)

References

  1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html

Role details: reporting, supervision and training

Each role links to its own terms of reference, which holds the full reporting line, supervision, training and competency record.

Role Reports to Supervises Core training
Manager ISCO 1342 The chief executive officer or the clinic owner. Nurses and the secretary in the laboratory. Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Nurse ISCO 2221 The manager. No direct reports. Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
Secretary ISCO 3344 The manager. No direct reports. Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)

Who is responsible for Standard 28

Each criterion below names the positions responsible for it.

Criterion Positions responsible or involved
28.1 Calibration Current and Traceable to a Reference Manager · Secretary
28.2 Reagent Lots Verified Before Use Nurse · Manager
28.3 Method Validated Before Clinical Use Manager · Secretary
28.4 Internal Quality Control Run Before Results Are Released Manager · Secretary
28.5 External Quality Assessment Participation Manager · Secretary
28.6 Test Indications, Limitations and Patient Preparation Given to Users Nurse · Manager
28.7 Specimens Accepted, Logged and Rejected by Defined Criteria Manager · Secretary
28.8 Specimen Transport Preserves Integrity Manager · Nurse
28.9 STAT Specimens Prioritised and Turnaround Monitored Manager · Secretary
28.10 Chain of Custody for Forensic or Legal Specimens Manager · Nurse
28.11 Critical Values Communicated and Receipt Verified Manager · Secretary
28.12 Results Authorised Before Release Manager · Nurse
28.13 Reports Meet a Defined Minimum and Reach the Correct Recipient Manager
28.14 Result Amendments Traceable to the Original Manager · Secretary
28.15 Suppliers and Reagents Qualified Manager · Secretary

Quality oversight

The quality chain keeps this standard under review, from the governing body down to the teams that run each process.

    Questions and answers

    The criterion text is the authority. These answers summarise it.

    Who does Standard 28 apply to?

    It applies to medical laboratories. It is required by default in hospitals, ambulatory clinics and primary health clinics, optional in other facility types that provide laboratory testing, and not applicable to telemedicine and fitness and wellness. Section 4 shows the detail.

    What do the colours mean for this standard?

    Red means Must, required in every setting. Yellow means Medium, required and adapted to what a setting can realistically do. Green means Minimum, the lowest acceptable level.

    Must every instrument be calibrated?

    Yes. Each instrument and measuring device is calibrated or verified at its set interval against a traceable reference (criterion 28.1).

    Must each new reagent lot be checked?

    Yes. Each new lot is checked against the previous lot or the acceptance criteria before it is used on patients (criterion 28.2).

    Must every method be validated?

    Yes. Each method is validated for its clinical use and revalidated after a significant change (criterion 28.3).

    Is internal quality control required for every run?

    Yes. Control material is run at the set interval and with each analytical run, and failures are acted on before results are released (criterion 28.4).

    Must the laboratory join external quality assessment?

    Yes, for each test where a scheme exists (criterion 28.5).

    Must clinicians receive test information?

    Yes. Clinicians and patients receive the indications, limitations and preparation for each test (criterion 28.6).

    When is a specimen rejected?

    When it does not meet the defined acceptance criteria. Every rejection is recorded with its reason and the clinician is told (criterion 28.7).

    Must specimens be transported under controlled conditions?

    Yes. Transport preserves integrity, and temperature and time are recorded for each shipment (criterion 28.8).

    What is a chain of custody?

    The documented record of every person who handled a forensic or legal specimen, from collection to disposal (criterion 28.10).

    What must happen to a critical value?

    It is communicated to a responsible clinician within the defined time, and receipt is recorded (criterion 28.11).

    Who may authorise a result?

    A qualified person, before the result is released. The authorisation is recorded (criterion 28.12).

    What must a report contain?

    The defined minimum information, including identifiers, test name, result, units, reference interval and the date and time (criterion 28.13).

    How is a change to a released result handled?

    It is recorded as an amendment, linked to the original result, with the reason given (criterion 28.14).

    Must suppliers be qualified?

    Yes. Suppliers of reagents, consumables and equipment services are qualified before use and reviewed at a set interval (criterion 28.15).

    Who is responsible for each criterion?

    The criteria assign responsibility to the manager, the nurse and the secretary. Each record should show who wrote and who reviewed it.

    Which e-learning courses are required?

    All fifteen criteria require the manager to complete Governance & Document Control for Laboratory Accreditation (LAB-01). Each criterion also names its own course. The full list is in section 11.

    Are the e-learning courses compulsory?

    Only the courses in the e-learning table marked as must take are required, for the roles named. Other courses are recommended.

    How is evidence shown to the surveyor?

    The records named in each criterion’s documents table, covering the period stated in the criterion.

    Do I have to use the forms on this page?

    No. They are reference drafts. Adapt them, write your own, or upload your own, provided the same criterion is met and the evidence can be shown.

    Does this standard set requirements for AI?

    No. None of its criteria sets a requirement for artificial intelligence. The criteria cover equipment, methods, specimens and reports directly.

    The criterion text is the authority. Do these answers replace it?

    No. These answers summarise the criterion. The criterion text is the authority.

    E-learning by standard

    These are the GMJ Academy courses named in the criteria of this standard. Each course name opens the course page in the academy. Where a course is published in English and in Georgian, both versions are linked. Evidence of completion is a verified micro-credential in the GMJ Academy micro-credential portfolio, or a training certificate in the staff file.

    E-learning course Who must take Who is recommended Criteria
    Governance & Document Control for Laboratory Accreditation (LAB-01) (LAB-01-EN)
    GMJ Academy: English · Georgian
    Manager Nurse, Secretary 28.1, 28.2, 28.3, 28.4, 28.5, 28.6, 28.7, 28.8, 28.9, 28.10, 28.11, 28.12, 28.13, 28.14, 28.15
    Equipment, Reagents & Traceability (LAB-04) (LAB-04-EN)
    GMJ Academy: English · Georgian
    Manager, Nurse None 28.1, 28.2, 28.15
    Personnel & Competence in Laboratory Accreditation (LAB-03) (LAB-03-EN)
    GMJ Academy: English · Georgian
    Manager, Nurse None 28.3, 28.4
    Facilities, Environment & Suppliers (LAB-05) (LAB-05-EN)
    GMJ Academy: English · Georgian
    Manager, Nurse None 28.5, 28.7, 28.8
    Post-Examination: Reporting & Critical Values (LAB-08) (LAB-08-EN)
    GMJ Academy: English · Georgian
    Manager, Nurse None 28.6, 28.9, 28.10, 28.11, 28.12, 28.13, 28.14

    Related courses in the GMJ Academy

    These courses cover the same subjects in more depth or for a specific type of facility. They are optional. They are not required by the criteria.

    Course Subject Criteria
    Equipment, Reagents & Traceability (LAB-04)
    GMJ Academy: English · Georgian
    Equipment calibration and traceability 28.1
    Facilities, Environment & Suppliers (LAB-05)
    GMJ Academy: English · Georgian
    Specimen acceptance and facilities 28.7
    Post-Examination: Reporting & Critical Values (LAB-08)
    GMJ Academy: English · Georgian
    Critical values and reporting 28.11

    Back to ASF Health Standards V11 · Standards Map · E-learning by standard · References

    References and guidelines for this standard

    The guidelines and references cited in the criteria above are listed here. Each link opens the full text.

    1. International Organization for Standardization (2022). ISO 15189:2022 Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
    2. International Organization for Standardization (2020). ISO 17511:2020 In vitro diagnostic medical devices: Metrological traceability of values assigned to calibrators and control materials. Geneva: ISO. https://www.iso.org/standard/69984.html
    3. World Health Organization. WHO operational handbook on tuberculosis: Module 3: Diagnosis. Geneva: WHO. https://www.who.int/publications/i/item/9789240110991
    4. World Health Organization. Manual for organizing a national external quality assessment programme for health laboratories and other testing sites. Geneva: WHO. https://www.who.int/publications/i/item/9789241549677
    5. World Health Organization (2020). Laboratory biosafety manual, 4th edition. Geneva: WHO. https://www.who.int/publications/i/item/9789240011311
    6. World Health Organization (2021). Global Patient Safety Action Plan 2021–2030: Towards eliminating avoidable harm in health care. Geneva: WHO. https://www.who.int/publications-detail-redirect/9789240032705

    Annex A. ISO 15189:2022 clause map

    Annexes A to D are part of Standard 28. They do not add criteria and do not change any rating. They show how the fifteen criteria relate to ISO 15189:2022, Medical laboratories: Requirements for quality and competence, and to the ISO 15189 Ready documents, and they list the additional criteria from the ISO 15189 module for a decision. ISO 15189:2022 is a copyrighted publication. This standard uses its clause titles only and does not reproduce its text.

    Each row names an ISO 15189:2022 clause area, the Standard 28 criteria that cover it, the ISO 15189 Ready documents that support it, and the status.

    Clause Area Subject Standard 28 criteria ISO 15189 Ready documents Status
    Clause 4 General requirements Impartiality, confidentiality and requirements regarding patients Not in the 15 core criteria (see Annex C: 29.1, 29.2)
    Not in the 15 core criteria (see Annex C: 29.3)
    Partly: 28.6 (test information)
    LAB-P-24, LAB-FR-RISK-04, LAB-SOP-HR-04; LAB-P-10; LAB-SOP-PA-01 Gap / partly covered
    Clause 5 Structural requirements Accountable person, authority, risk management (trace) None (Annex C: 29.4, 29.6)
    None (Annex C: 29.5)
    LAB-P-01, LAB-JD-01, LAB-P-04, LAB-PR-03, LAB-R-10 Gap
    Clause 6 Resource requirements: personnel Qualification, competence, training, staffing (trace) None (Annex C: 29.8 to 29.12) LAB-P-08, LAB-P-09, LAB-PR-07, LAB-PR-08, LAB-SOP-HR-01 to HR-03 Gap
    Clause 6 Resource requirements: equipment Calibration, traceability, maintenance, malfunction (trace) 28.1 Calibration and traceability
    None for maintenance (Annex C: 29.14, 29.17)
    LAB-P-18, LAB-P-19, LAB-PR-10, LAB-PR-11, LAB-SOP-A-05, LAB-SOP-EQ-01 to EQ-04 Covered for calibration; maintenance is Annex
    Clause 6 Resource requirements: reagents and consumables Reagent lot verification (trace) 28.2 Reagent lot verification LAB-SOP-A-09, LAB-SOP-QC-02, LAB-SOP-BIO-06, LAB-SOP-IMM-04 Covered
    Clause 6 Resource requirements: externally provided products and services Suppliers and services (trace) 28.15 Suppliers qualified and reviewed LAB-P-22, LAB-PR-29, LAB-SOP-CUST-04 Covered
    Clause 6 Resource requirements: facilities, environment, safety, utilities, waste Accommodation, environment, PPE, waste, utilities (trace) None (Annex C: 29.18 to 29.23, 29.25) LAB-P-12, LAB-P-14, LAB-P-21, LAB-P-28, LAB-PR-12, LAB-PR-13, LAB-SOP-SF-01 to SF-07 Gap
    Clause 6 / 7.6 Information management Information systems, access, data security (trace) None (Annex C: 29.20, 29.26) LAB-P-11, LAB-P-43, LAB-SOP-31, LAB-SOP-IT-01, LAB-FR-IT-01 to IT-06 Gap
    Clause 7.2 Pre-examination processes Request, patient preparation, specimen acceptance, transport, storage, chain of custody 28.6 Test information
    28.7 Specimen acceptance
    28.8 Transport
    28.9 STAT and turnaround (partly)
    28.10 Chain of custody
    LAB-P-16, LAB-P-17, LAB-PR-17, LAB-PR-18, LAB-SOP-PA-01 to PA-12, LAB-FR-SM-05 to SM-08 Covered in part; identification and storage are Annex
    Clause 7.3 Examination processes: method validation Method verification and validation before clinical use (trace) 28.3 Method validation LAB-P-36, LAB-SOP-A-01, LAB-SOP-A-02, LAB-SOP-BIO-08 Covered
    Clause 7.3 Examination processes: quality control Internal quality control, rules, runs (trace) 28.4 Internal quality control LAB-P-37, LAB-PR-23, LAB-SOP-A-03, LAB-SOP-A-06, LAB-SOP-QC-01 Covered
    Clause 7.3 Examination processes: external quality assessment EQA participation and follow-up (trace) 28.5 External quality assessment LAB-SOP-A-04, LAB-SOP-23, LAB-SOP-BIO-04, LAB-SOP-COAG-04, LAB-SOP-HEM-06, LAB-SOP-IMM-03 Covered
    Clause 7.3 Examination processes: measurement uncertainty Estimation and reporting of measurement uncertainty (trace) None in the 15 core criteria LAB-SOP-40-MU, LAB-FR-MU-01 Gap (trace numbering to check)
    Clause 7.4 Post-examination processes: validation, authorisation and release Result review, authorisation, reporting (trace) 28.12 Result authorisation
    28.13 Report content
    Recipient check: Annex C 29.42
    LAB-P-39, LAB-P-42, LAB-SOP-18, LAB-SOP-PO-01, LAB-SOP-PO-02, LAB-PR-22, LAB-PR-32, LAB-MX-01 Covered in part
    Clause 7.4 Post-examination processes: critical values Critical value communication and receipt (trace) 28.11 Critical values LAB-P-40, LAB-SOP-19, LAB-SOP-PO-03, LAB-SOP-COAG-06, LAB-SOP-HEM-07 Covered
    Clause 7.4 Post-examination processes: amendments Traceable correction of released results (trace) 28.14 Amendments LAB-SOP-PO-07, LAB-PR-32, LAB-P-42 Covered
    Clause 7.4 Post-examination processes: turnaround, referrals, comments Turnaround, referred tests, cumulative results, interpretive comments (trace) 28.9 Turnaround (partly)
    Others: Annex C 29.44 to 29.46
    LAB-P-38, LAB-SOP-CUST-03, LAB-SOP-IMM-05 Partly covered
    Clause 8 Management system: documents and records Document control, approval, retention (trace) None (Annex C: 29.48, 29.49, 29.50) LAB-P-02, LAB-P-03, LAB-PR-01, LAB-PR-02, LAB-R-01, LAB-R-02, LAB-MX-04, LAB-FR-DOC-01 Gap
    Clause 8 Management system: internal audit, nonconformity, corrective and preventive action Audit, corrective action, prevention (trace: 8.7, 8.8) None (Annex C: 29.51 to 29.53) LAB-P-05, LAB-P-06, LAB-PR-04, LAB-PR-05, LAB-SOP-21, LAB-R-07 Gap
    Clause 8 Management system: management review Review, minutes, actions (trace: 8.9) None (Annex C: 29.54) LAB-P-07, LAB-PR-06, LAB-FR-MRM-04, LAB-FR-MRM-05, LAB-FR-MRM-06 Gap
    Clause 8 Management system: complaints, feedback and improvement Complaints, user feedback, improvement (trace) None (Annex C: 29.55 to 29.57) LAB-P-23, LAB-SOP-CUST-01, LAB-SOP-CUST-02, LAB-FR-CUST-05, LAB-P-25 Gap

    Subclause numbers marked (trace) come from the internal source trace and are to be confirmed against the licensed ISO 15189:2022 text.

    Annex B. ISO 15189 Ready documents by criterion

    Each core criterion lists the ISO 15189 Ready documents that support it. Codes are from the ISO 15189 Ready master index.

    Georgian edition of the ISO 15189 Ready documents: ქართული (Georgian).

    Criterion Topic ISO 15189 Ready documents (code)
    28.1 Calibration and traceability LAB-P-19, LAB-PR-10, LAB-SOP-A-05, LAB-SOP-EQ-02
    28.2 Reagent lot verification LAB-SOP-A-09, LAB-SOP-QC-02, LAB-SOP-BIO-06, LAB-SOP-IMM-04
    28.3 Method validation LAB-P-36, LAB-SOP-A-01, LAB-SOP-A-02, LAB-SOP-BIO-08
    28.4 Internal quality control LAB-P-37, LAB-PR-23, LAB-SOP-A-03, LAB-SOP-QC-01, LAB-SOP-BIO-03, LAB-SOP-COAG-03, LAB-SOP-HEM-03, LAB-SOP-IMM-02, LAB-SOP-UR-03
    28.5 External quality assessment LAB-SOP-A-04, LAB-SOP-23, LAB-SOP-BIO-04, LAB-SOP-COAG-04, LAB-SOP-HEM-06, LAB-SOP-IMM-03
    28.6 Test information for users LAB-SOP-PA-01, LAB-P-25, LAB-SOP-PO-06
    28.7 Specimen acceptance and rejection LAB-P-17, LAB-PR-17, LAB-PR-18, LAB-SOP-PA-06, LAB-SOP-PA-07, LAB-SOP-UR-04, LAB-SOP-COAG-05
    28.8 Specimen transport LAB-SOP-PA-04, LAB-SOP-PA-05, LAB-P-16
    28.9 STAT and turnaround LAB-P-38, LAB-PR-22
    28.10 Chain of custody LAB-SOP-PA-12, LAB-FR-SM-08
    28.11 Critical values LAB-P-40, LAB-SOP-19, LAB-SOP-PO-03, LAB-SOP-COAG-06, LAB-SOP-HEM-07
    28.12 Result authorisation LAB-P-42, LAB-SOP-18, LAB-PR-32, LAB-MX-01
    28.13 Report content LAB-P-39, LAB-SOP-PO-01, LAB-SOP-PO-02
    28.14 Amendments LAB-SOP-PO-07, LAB-PR-32, LAB-P-42
    28.15 Suppliers and external services LAB-P-22, LAB-PR-29, LAB-SOP-CUST-04

    Annex C. Additional criteria from the ISO 15189 module, for decision

    These are the 57 criteria from the laboratory module (version 4). Some are already covered by a core criterion. The others are listed here for a decision on whether to adopt them as additional criteria. Adopting them would change the standard and needs approval.

    Module ID Criterion (module wording) ISO 15189 clause group Core criterion Status
    29.1 Impartiality: No Commercial Pressure on Results 4 Annex only Not in the 15 core criteria
    29.2 Conflict of Interest Declared and Reviewed 4 Annex only Not in the 15 core criteria
    29.3 Patient Information Is Kept Confidential 4 Annex only Not in the 15 core criteria
    29.4 Named Accountable Person 5 Annex only Not in the 15 core criteria
    29.5 Risk Register, Reviewed on a Schedule 5 Annex only Not in the 15 core criteria
    29.6 Authority and Responsibilities Are Defined and Documented 5 Annex only Not in the 15 core criteria
    29.7 External Service and Supply Evaluation 6 28.15 (partly) Covered by core criterion
    29.8 Documented Qualification Before Independent Work 6 Annex only Not in the 15 core criteria
    29.9 Competence Reassessed on a Schedule 6 Annex only Not in the 15 core criteria
    29.10 Orientation Before First Independent Shift 6 Annex only Not in the 15 core criteria
    29.11 Training Records Tied to Specific SOPs 6 Annex only Not in the 15 core criteria
    29.12 Adequate Staffing for Safe Operation 6 Annex only Not in the 15 core criteria
    29.13 Calibration Current on Every Reportable Instrument 6 28.1 Covered by core criterion
    29.14 Preventive Maintenance on a Fixed Schedule 6 Annex only Not in the 15 core criteria
    29.15 Reagent Lot Verification Before Use 6 28.2 Covered by core criterion
    29.16 Metrological Traceability to a Recognized Reference 6 28.1 Covered by core criterion
    29.17 Equipment Malfunction: Documented Response 6 Annex only Not in the 15 core criteria
    29.18 Personal Protective Equipment, Available and Used 6 Annex only Not in the 15 core criteria
    29.19 Exposure Incident: Documented Response Protocol 6 Annex only Not in the 15 core criteria
    29.20 Information System Access Controlled and Logged 6 Annex only Not in the 15 core criteria
    29.21 Laboratory Accommodation Protects Specimens and Staff 6 Annex only Not in the 15 core criteria
    29.22 Environmental Conditions Are Monitored and Recorded 6 Annex only Not in the 15 core criteria
    29.23 Laboratory Waste Is Segregated and Disposed of Safely 6 Annex only Not in the 15 core criteria
    29.24 Suppliers and External Services Are Qualified and Monitored 6 28.15 Covered by core criterion
    29.25 Laboratory Utilities and Failures Are Planned For 6 Annex only Not in the 15 core criteria
    29.26 Laboratory Information Is Managed Securely and Accurately 6 / 7.6 Annex only Not in the 15 core criteria
    29.27 Two-Identifier Patient Verification 7.2 Annex only Not in the 15 core criteria
    29.28 Specimen Labeled at the Point of Collection 7.2 Annex only Not in the 15 core criteria
    29.29 Rejection Criteria, Applied Consistently 7.2 28.7 Covered by core criterion
    29.30 Transport Conditions Preserve Specimen Integrity 7.2 28.8 Covered by core criterion
    29.31 Chain of Custody for Forensic or Legal Specimens 7.2 28.10 Covered by core criterion
    29.32 Specimen Accessioning Logged in Real Time 7.2 Annex only Not in the 15 core criteria
    29.33 Urgent/STAT Specimens Prioritized and Tracked 7.2 28.9 Covered by core criterion
    29.34 Specimen Retention Before Discard 7.2 Annex only Not in the 15 core criteria
    29.35 Method Validation Before Clinical Use 7.3 28.3 Covered by core criterion
    29.36 Internal Quality Control Run Before Patient Results Released 7.3 28.4 Covered by core criterion
    29.37 External Quality Assessment Participation 7.3 28.5 Covered by core criterion
    29.38 Critical Value Communication, Verified Received 7.4 28.11 Covered by core criterion
    29.39 Result Amendment Traceable to Original 7.4 28.14 Covered by core criterion
    29.40 Result Authorization Before Release 7.4 28.12 Covered by core criterion
    29.41 Report Content Meets a Defined Minimum 7.4 28.13 Covered by core criterion
    29.42 Report Delivered to the Correct Recipient 7.4 Annex only Not in the 15 core criteria
    29.43 Turnaround Time Monitored Against a Target 7.4 28.9 Covered by core criterion
    29.44 Supplementary and Referred Test Results Tracked to Closure 7.4 Annex only Not in the 15 core criteria
    29.45 Cumulative/Historical Results Available at Review 7.4 Annex only Not in the 15 core criteria
    29.46 Interpretive Comments Added Where Clinically Needed 7.4 Annex only Not in the 15 core criteria
    29.47 Patients Are Informed How to Prepare for Testing 7.2 / 4.3 28.6 Covered by core criterion
    29.48 Document Control: One Current Version 8 Annex only Not in the 15 core criteria
    29.49 Document Approval Before Use 8 Annex only Not in the 15 core criteria
    29.50 Retention Schedule Matches Legal Minimum 8 Annex only Not in the 15 core criteria
    29.51 Internal Audit, On a Fixed Schedule 8 Annex only Not in the 15 core criteria
    29.52 Corrective Action Closes the Root Cause 8 Annex only Not in the 15 core criteria
    29.53 Preventive Action, Not Only Reactive 8 Annex only Not in the 15 core criteria
    29.54 Management Review, With Minutes and Actions 8 Annex only Not in the 15 core criteria
    29.55 Complaints Logged, Investigated, Closed 8 Annex only Not in the 15 core criteria
    29.56 User Feedback Is Gathered and Acted On 8 Annex only Not in the 15 core criteria
    29.57 Improvement Is Planned, Measured and Reviewed 8 Annex only Not in the 15 core criteria

    Source trace: the trace file groups criteria by clause and gives subclause numbers from published summaries. Those subclause numbers are marked (trace) and must be checked against the licensed ISO 15189:2022 text before publication.

    Numbering mismatch: the trace lists criteria 29.50 to 29.59 and assigns criterion 29.49 to measurement uncertainty. The module (v4) has 57 criteria, ending at 29.57, and criterion 29.49 is Document Approval. Measurement uncertainty has no criterion in the module and is listed as a gap in Annex A.

    Master index: the 184 ISO 15189 Ready documents are listed by category and code in the master index. Category ranges are not clause-mapped in the index; Annex B maps the core criteria to document codes.

    Annex D. ISO 15189 references

    The primary reference is the ISO standard itself. The Eurachem leaflet summarises the 2022 changes. Both are listed for the assessor to check against the licensed text.

    1. ISO 15189:2022. Medical laboratories: Requirements for quality and competence. Geneva: ISO. https://www.iso.org/standard/76677.html
    2. Eurachem (2022). Leaflet: ISO 15189:2022 (changes from the 2012 edition). Eurachem. https://eurachem.org/images/stories/leaflets/15189_2022/Eurachem_Leaflet_15189_2022_EN.pdf

    Annex E. Coverage of ISO 15189 across all ASF standards

    Annex E shows where each ISO 15189:2022 clause area is covered across all ASF Health Standards, not only in Standard 28. The aim is full coverage of ISO 15189:2022. At the time of this draft, some areas are covered only in part, and some have no criterion yet. Each gap is named with the action that closes it. Full coverage can be claimed only when every row is marked Covered.

    Of the 21 ISO 15189:2022 clause areas, 8 are covered by ASF criteria, 9 are partly covered, and 4 have no criterion yet. Accreditation covers 100% of ISO 15189 only when every row below is Covered.

    ISO 15189 clause area Where ASF covers it (live Version 11 criteria) Status Action to reach full coverage
    Impartiality and freedom from commercial pressure (4.1) No criterion in any standard Not yet covered Add a criterion to Standard 2 (governance): clinical decisions are free from commercial and financial pressure, and conflicts of interest are declared and reviewed.
    Confidentiality of patient information (4.2) Standard 2, 2.4 Patient Data Is Kept Secure; 2.5 Records Are Kept Exactly as Long as Required Partly Add a laboratory confidentiality criterion (Annex C 29.3) to Standard 28, or extend 2.4 to cover test results.
    Requirements regarding patients (4.3) Standard 28, 28.6 Test Indications, Limitations and Patient Preparation; Standard 2, 2.10 Patients and Families Have a Voice Covered None beyond the Annex C criteria on patient identification (29.27, 29.28).
    Accountable person and authority (5.4) Standard 2, 2.1 The Board Is Accountable Partly Add a named laboratory accountable person criterion (Annex C 29.4 and 29.6) to Standard 28.
    Risk management (5.6) Standard 3, 3.6 Facility Risks Are Tracked in One Integrated Register (facility level) Partly Confirm that 3.6 includes laboratory risks, or add a laboratory risk register (Annex C 29.5).
    Personnel: qualification, competence, training, staffing (6) Standard 1, 1.1 Staffing Matches Patient Need; Standard 6, 6.1 Ongoing Competency Is Regularly Evaluated, 6.2 Ongoing Education, 6.10 Staff Can Raise Concerns Covered Laboratory-specific training records (Annex C 29.8 to 29.11) are optional; the core criteria cover the requirement.
    Equipment, calibration and traceability (6) Standard 28, 28.1 Calibration Current and Traceable to a Reference; Standard 3, 3.4 Medical Equipment Is Maintained on Schedule Covered None.
    Reagents and consumables (6) Standard 28, 28.2 Reagent Lots Verified Before Use Covered None.
    External services and suppliers (6) Standard 28, 28.15 Suppliers and Reagents Qualified Covered None.
    Facilities, environment, safety and waste (6) Standard 3, 3.1 to 3.17 (premises, water, cleanliness, equipment); no core criterion for laboratory waste or biosafety Partly Add a laboratory biosafety and waste criterion (Annex C 29.22, 29.23) to Standard 28 or to Standard 3.
    Information systems and data (6, 7.6) Standard 2, 2.4 Patient Data Is Kept Secure; Standard 5, 5.1 Digital Care Systems Follow an Assessment and Management Process, 5.3 Technical Expertise Partly Add laboratory information system access, audit trail and backup criterion (Annex C 29.20, 29.26).
    Pre-examination processes (7.2) Standard 28, 28.6 to 28.10 (information, acceptance, transport, STAT, chain of custody) Covered Identification and labelling (Annex C 29.27, 29.28) are optional additions.
    Examination: method validation, quality control, EQA (7.3) Standard 28, 28.3 Method Validated Before Clinical Use; 28.4 Internal Quality Control; 28.5 External Quality Assessment Covered None.
    Measurement uncertainty (7.3) No criterion in any standard Not yet covered Add a measurement uncertainty criterion to Standard 28.
    Post-examination processes: authorisation, critical values, reports, amendments (7.4) Standard 28, 28.11 Critical Values; 28.12 Result Authorisation; 28.13 Report Content; 28.14 Amendments Covered Recipient check (Annex C 29.42) is optional.
    Document control and records (8.2, 8.3) No explicit document control criterion; Standard 2, 2.3 Policy Gets Followed Not yet covered Add a document control criterion (one current version, approval before use) to Standard 2.
    Internal audit (8.8) No criterion in any standard; Standard 63, 63.5 audits handover only Not yet covered Add an internal audit criterion to Standard 2.
    Nonconformity, corrective and preventive action (8.7) Standard 2, 2.6 Serious Incidents Get Properly Investigated (incidents only) Partly Add a corrective and preventive action criterion (Annex C 29.52, 29.53) to Standard 2.
    Management review (8.9) Standard 2, 2.2 Written Strategic Plan; Standard 18, 18.6 Site-Level Quality Data Is Reviewed by Central Governance Partly Add a management review criterion with minutes and actions (Annex C 29.54) to Standard 2.
    Complaints and user feedback (8.6 or equivalent) Standard 2, 2.8 Patient Experience Is Measured Continuously; Standard 6, 6.6 Workforce Feedback Is Gathered and Acted On Partly Add a complaints process criterion (Annex C 29.55, 29.56) to Standard 2.
    Improvement (8.5) Standard 2, 2.7 Safety Culture Is Measured Partly Add an improvement planning criterion (Annex C 29.57) to Standard 2.

    Criterion numbers are checked against the live Version 11 overview. Module 29 IDs are checked against the laboratory module (version 4). Subclause numbers are to be confirmed against the licensed ISO 15189:2022 text.